Always take into account the markings provided on the inside and outside of
the printer. A brief overview of these markings and their meaning is given
below.
Symbol
Figure 14: Caution hot:
Figure 15: High voltage
Figure 16: Type B
equipment:
Figure 17: Supplementa
ry protective earth
connector:
Figure 18: Intergroundi
ng connector:
Figure 19: Protective
earth (ground):
Figure 20: Power button
2852 A EN 20070205
Explanation
Keep hands clear from the thermal print head.
In order to reduce the risk of electric shock, do not
remove any covers.
Indicates that the Drystar AXYS complies with the lim-
its for type B equipment.
Provides a connection between the Drystar AXYS and
the potential equalization bus bar of the electrical sys-
tem as found in medical environments. This plug
should never be unplugged before the power is turned
off and the power plug has been removed.
Provides a connection between the printer and other
equipment, which might exhibit minor ground poten-
tial differences. These differences may degrade the
quality of communication between different equip-
ment. Never remove connections to this terminal.
Provides a connection between the printer and the
protective earth of the mains. Do not remove this con-
nection, because this will have a negative influence on
the leakage current.
Note that the power cord has to be disconnected from
the wall outlet in order to disconnect the unit entirely
from the mains.