Serial interface

At the rear of each machine a 9 pole
Sub-D plug for data transfer between
the Profitronic system and an external
report printer or PC is supplied.
The serial interface is RS 232
compatible.
For the interface configurations see
"PC/Printer Function" in the
Programming Manual.
Various printers can be used as
external printers:
– Epson-compatible
(contact Miele for a list of suitable
models)
– HP LaserJet
The address of the 9-pole pin sub-D
plug of the serial interface cable:
5GND (ground)
3TXD (transmit)
2RXD (receive)
1-4-6 (linked)
7-8 (linked)
A standard null-modem or LapLink
cable can be connected.
The extension cable to the printer/PC
must not exceed 42 ½ ft.(13 m).
Please note the following when
connecting a printer or PC:
– Only use an industry-standard PC or
printer.
– The size of the printer or PC must be
taken into account when installing
the machine.
32
The report printer settings are
described in detail in the Programming
manual under "PC/Printer function".
The following features can be printed:
a Programs
b Additional functions
c Rinse protocol
d Fault protocol