Interrupting A Program - Miele G 7836 CD Operating And Installation Instructions

Laboratory glassware washer
Hide thumbs Also See for G 7836 CD:
Table of Contents

Advertisement

Interrupting a program

A program can only be interrupted in
the operating levels B or D.
If the door needs to be opened in
extreme cases, e.g. if items are moving:
When continuing an interrupted
DESIN (disinfection) program, check
the message in the display at the
end of the program. If the message
PROCESS PARAMETERS NOT MET
is displayed, the door was opened
after the disinfection sequence of
the program and the DESIN
parameters were not fulfilled. The
program must be repeated.
In operating levels B or D
^ Press the "Stop" 7 button, the
program will be interrupted.
The display shows >CANCEL< OR
>CONTINUE< at water temperatures
below 105°F (40°C). It shows
>CANCEL< at water temperatures of
105°F (40°C) or higher.
^ Open the door.
,
Caution! Water and items in the
machine may be hot. Danger of
scalding or burning.
During programs with chemical
disinfection, hot steam with high
contents of disinfection agents could
occur.
^ Rearrange the load. Follow infection
control regulations and wear
protective gloves.
^ Carefully close the door.
,
If there is a lot of hot water in the
wash cabinet and the door is closed
quickly, hot water may spray out.
Danger of scalding/burning.
If the water temperature was below
105°F (40°C) when the program was
interrupted:
^ Press the "Start" 6 button.
The program continues.
If the water temperature was above
105°F (40°C) when the program was
interrupted:
^ Press the "Start" 6 button.
The program is canceled and the water
is drained. DRAIN WATER appears in
the display.
After the water was drained the
program can be restarted.