Range Calibration (P650 to P654)

There are two possible types of calibration:
Offset
Sound Velocity
Do Offset calibration at any steady level, unless a Sound Velocity calibration is also done.
If both calibrations are done, then do Offset at a known high level, and Sound Velocity at
a known low level.
7ML19985HT01
Index 2
00000 to 99999
Range:
15
twice to focus control on the secondary index.
twice and use the ARROW keys or enter the value for the next
Adjusts the measurement by a fixed amount
Adjusts speed of sound, and changes the measurement
calculations