Honeywell STT25D Operator's Manual page 23

Stt 3000-series stt250 smart temperature transmitter
Table of Contents

Advertisement

SYMPTOM
High input
Unstable
onscale input
Fail-safe
output signal
20
Smart Temperature Transmitter STT250 – Operator Manual
SFC MESSAGE/
DISPLAY
"INPUT OPEN" (with
upscale fail-safe)
"I/P OUT OF SPEC"
(indicates that the
input is above URL or
below LRL)
"STATUS CHECK =
O.K." since any
identified problem
would give upscale or
downscale fail-safe.
"CRITICAL STATUS" A fail-safe
POSSIBLE
CAUSE
Bad sensor wire
Check field wiring and
connection.
connections for partial
short circuit.
Check STT250
connections are correct
polarity.
Incorrect
Check that the
operating
transmitter is in its
voltage.
voltage operating area
and line resistance is
not excessive.
Bad sensor wire
Check connection and
connection.
wiring for intermittent
connections.
Check that sensor fault
detection is "ON", this
Intermittent
allows detection of a
open circuit of
bad sensor.
sensor.
Protect wiring by using
appropriate grounding,
Wiring disturbed
shielding etc.
by strong
electromagnetic
interference.
The SFC will indicate
output signal
the source of the
(critical status)
problem by displaying
can be caused
the appropriate error
by several
message.
reasons.
CURE

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Stt25hStt25m

Table of Contents