Fluke 1529 Technical Manual page 120

Chub-e4 thermometer
Hide thumbs Also See for 1529:
Table of Contents

Advertisement

1529 Chub-E4 Thermometer Readout
Commands
7.4.12.6
*SRE <num>|MIN|MAX|DEF
Sets the Service Request Enable Register.
*SRE 32
7.4.12.7
*STB?
Returns the Status Byte Register.
*STB?
200
7.4.12.8
*TST?
Performs a self-test and reports any errors that are found.
*TST?
0
7.4.12.9
STATus:MEASure?
Reads and clears the Measurement Status Register.
STAT:MEAS?
1
The assignment of the bits in this register is as follows:
bit 0: channel 1 measurement complete
bit 1: channel 2 measurement complete
bit 2: channel 3 measurement complete
bit 3: channel 4 measurement complete
bits 4-15: not used
7.4.12.10
STATus:MEASure:CONDition?
Returns the Measurement Status Condition Register.
STAT:MEAS:COND?
1
The assignment of the bits in this register is as follows:
bit 0: channel 1 measurement in process
110