● Do not use this control solution beyond the expiry date or
3 months after first opening. Write the opening date on the
control solution vial and discard the remaining solution after 3
months.
● It is recommended that the glucose strips control solution test
be done at room temperature (20°C to 25°C / 68°F to 77°F).
Make sure your control solution, device, and test strips are at
this specified temperature range before testing.
● Shake the bottle before use, discard the first drop of control
solution, and wipe off the dispenser tip to ensure a pure
sample and an accurate result.
● Store the control solution tightly closed at temperatures
between 2°C to 30°C (35.6°F to 86°F). Do NOT freeze.

SYSTEM TROUBLESHOOTING

If you follow the recommended action but the problem persists,
or error messages other than the ones below appear, please
call GlucoRx Customer care on 01483 755133. Do not attempt
to ever repair or disassemble the device yourself.

Result Readings (for glucose test)

MESSAGE
(warning)
Your blood glucose level is below 0.5 mmol/L
Your blood glucose level is 15 mmol/L or more
Your blood glucose level is above 38.9 mmol/L
WHAT IT MEANS