Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below,
and should only be used in such environments:
Immunity test
mitter
to
Conducted RF
IEC 61000-4-6
z
√P
Radiated RF
IEC 61000-4-3
15-0150 BPW-710-CA_IM_v3--03.indd 32
IEC 60601
Compliance
test level
level
3 Vrms
150 kHz
to
Not
80 MHz
Applicable
3 V/m
3 V/m
80 MHz
to
2.5 GHz
Electromagnetic environment – guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the device, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W) ac-
cording to the transmitter manufacturer and
d is the recommended separation distance
in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,a should be less than the compli-
ance level in each frequency range.b
Interference may occur in the vicinity
of equipment marked with the following
symbol: