Agilent Technologies 490 User Manual
Agilent Technologies 490 User Manual

Agilent Technologies 490 User Manual

Micro gc biogas analyzer
Hide thumbs Also See for 490:

Advertisement

Quick Links

Agilent 490 
Micro GC Biogas
Analyzers
User Manual

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies 490

  • Page 1 Agilent 490  Micro GC Biogas Analyzers User Manual...
  • Page 2: Manual Part Number

    Further, to the max- agreement and written consent from  imum extent permitted by applicable Agilent Technologies, Inc. as governed by CAU TI O N law, Agilent disclaims all warranties, United States and international copyright either express or implied, with regard laws.
  • Page 3: The Agilent Biogas Analyzers

    The Agilent 490 Micro GC Biogas Analyzer is a factory tuned analyzer specific for biogas analysis. Two types of Biogas Analyzers are available: Agilent 490 Micro GC Biogas Analyzer and the Agilent 490 Micro GC Biogas Analyzer Extended.
  • Page 4: Checkout Information

    The Biogas Analyzer is factory tuned including appropriate settings for backflush times for the CP- Molsieve 5A and CP-PoraPLOT U channel. Final checkout of the Agilent 490 Micro GC Biogas Analyzers is performed with a Universal Gas Calibration standard. This contains helium, neon, hydrogen,...
  • Page 5: Cp-Molsieve Channel

    The Agilent Biogas Analyzers CP-Molsieve channel The CP-Molsieve 5A channel of the Biogas Analyzer analyzes permanent gases such as helium, neon, hydrogen, oxygen, nitrogen, methane and carbon monoxide. Figure 1 is a chromatogram of the Universal gas Calibration Standard analyzed on the CP-Molsieve 5A channel. Seconds Figure 1 The Universal Gas Calibration Standard analyzed on the CP-Molsieve channel...
  • Page 6: Cp-Poraplot U Channel

    The Agilent Biogas Analyzers CP-PoraPLOT U channel The CP-PoraPLOT U channel of the Biogas Analyzer is used to analyze carbon dioxide, ethane, hydrogen sulfide and n-propane. This channel and the sample inlet are Ultimetal deactivated which results in improved performance for hydrogen sulfide analysis.
  • Page 7 The Agilent Biogas Analyzers CP Sil 5 CB channel The CP-Sil 5 CB channel of the Biogas Analyzer Extended is used to analyze the hydrocarbon up to n-heptane. Hydrocarbons that are specified for this analyzer are i-butane, n-butane, neo-pentane, n-pentane, iso-pentane, n-hexane, and n-heptane. The Universal Gas Calibration Standard does not contain all these hydrocarbons.
  • Page 8: Tuning The Backflush Time

    The Agilent Biogas Analyzers Tuning the backflush time For the CP-Molsieve 5A and the CP-PoraPLOT U channel, the backflush times must be tuned. For tuning, the Universal Gas Calibration Standard is recommended. The purpose of tuning the CP-Molsieve 5A is to get all methane on the column while all components such as moisture, carbon dioxide, ethane and higher hydrocarbons are backflushed.
  • Page 9: Appendix 1: Certificate Of The Universal Gas Calibration Standard

    The Agilent Biogas Analyzers Appendix 1: Certificate of the Universal Gas Calibration Standard Part No: 5184-3541 Part No. Kit: 5184-3546 Biogas Analyzers...
  • Page 10: Appendix 2: The Biogas Analyzer Method

    The Agilent Biogas Analyzers Appendix 2: The Biogas Analyzer method Table 1 The Biogas Analyzer method Method Settings CP-Molsieve 5A CP-PoraPLOT U CP-Sil 5 CB Carrier gas Argon Helium Helium Injector Temperature (°C) Injection Time (ms) Backflush time (s) Column Temperature (°C) Pressure (kPa) Sample Line temperature (°C) Invert signal...
  • Page 12 INGENIERIA ANALITICA S.L. Tel: (+34) 902.45.6677 Fax: (+34) 902.46.6677 www.ingenieria-analitica.com inf@ingenieria-analitica.com © Agilent Technologies, Inc. Printed in USA, February 2012...