Quick Start

SITRANS LR 460 only supports SIMATIC PDM version 6.0 with SP2.
Note:
To set up SITRANS LR 460 for a simple application requires only the following settings:
select application type (silo construction)
select operation mode: level, distance or space
set speed of response
set High and Low Calibration Points
A Quick Start wizard groups together all the settings you need. There are two ways to
access the wizard:
Quick Start Wizard via the handheld programmer
Quick Start Wizard via SIMATIC PDM

Activating SITRANS LR 460

Notes:
• Keep infrared devices such as laptops, cell phones, and PDAs, away from
SITRANS LR 460 to prevent inadvertent operation.
• Frequently switching the device off and on causes wear of the electronics (see
Statistics
on page 48 to access the records).
Power up the instrument. SITRANS LR 460 automatically starts up in RUN mode, and
detects the distance to the material level referenced from the sensor flange face. The
LCD displays the measurement and the default unit is meters. System status is displayed
on the LCD, or on a remote communications terminal.

RUN mode display

Normal operation
98.82
3.800 mA
mA value, HART only (press key 5 on
handheld programmer to activate)
RUN mode (startup display)
See
1.
Loss of Echo (LOE)
See
2.
Flashes once per second to indicate that the device is functioning.
Page 26
reliable echo
level
indicator
units
m
on page 83 for more detailed information.
on page 72 for more detail.
SITRANS LR 460 – INSTRUCTION MANUAL
on page 28
on page 30
1 2
LOE condition
unreliable echo
1
indicator
flashing
2
98.82
heartbeat
S: 0
2.3.
1
m
fault code
7ML19985JM01