Fluke 1523 Technical Manual page 90

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

2. Set the UUT password enable by sending the following commands through the
3. Set the UUT to the L75_OHMS cal range by sending the following commands
4. Query the Ref points by sending the following Command:
5. Query the ADJ values for the range by sending the following Command:
6. Set the UUT in the LO_OHMS cal range by sending the following commands
7. Query the Ref points by sending the following Command:
8. Query the ADJ values for the range by sending the following Command:
9. Set the UUT in the MED_OHMS cal range by sending the following commands
10. Query the Ref points by sending the following Command:
11. Query the ADJ values for the range by sending the following Command:
12. Set the UUT in the HI_OHMS cal range by sending the following commands
13. Query the Ref points by sending the following Command:
14. Query the ADJ values for the range by sending the following Command:
Shop for Fluke products online at:
Calibration of Your Reference Thermometer Readout
serial port:
SYST:PASS:CEN XXXX (XXXX is the current UUT password)
Query the password enable to ensure communication is working and that it is
enabled before continuing. Do this by sending the following command:
SYST:PASS:CEN?
The returned value should be a 1. If not recheck the serial cable connection and
the terminal program communication settings listed in step 1.
through the serial port:
CAL1:DEV:RANG:SEL L75_OHMS
CAL1:DEV:RANG:REF?
CAL1:DEV:RANG:ADJ?
The ADJ values will be in order of the ref points queried in the previous step.
through the serial port:
CAL1:DEV:RANG:SEL LO_OHMS
CAL1:DEV:RANG:REF?
CAL1:DEV:RANG:ADJ?
The ADJ values will be in order of the ref points queried in the previous step.
through the serial port:
CAL1:DEV:RANG:SEL MED_OHMS
CAL1:DEV:RANG:REF?
CAL1:DEV:RANG:ADJ?
The ADJ values will be in order of the ref points queried in the previous step.
through the serial port:
CAL1:DEV:RANG:SEL HI_OHMS
CAL1:DEV:RANG:REF?
CAL1:DEV:RANG:ADJ?
MyFlukeStore
www.
Manual Calibration Process