1523, 1524 Reference Thermometer

SETUP

In this mode the data stored and displayed last remains on screen, this includes statis-
tics information. Press the HOLD or HOME keys to exit the hold mode.
4.5 SETUP
The SETUP key accesses the SETUP Menu and allows access to Channel T1 Setup
submenus, Channel T2 Setup submenus (1524 only), Instrument setup, and the Date/
Time (1524 only) submenus. The submenus allow the user to configure the instrument
and its system parameters.
4.5.1

Channel T1

Select the Channel T1 submenu by pressing the ENTER key. This menu allows the
user access to view the probe data from the INFO-CON connector and to setup the
auxiliary display for channel T1.
4.5.1.1

Probe

This is a read only field; it tells the user the probe and/or the conversion type used with
the related probe.
4.5.1.2

Config

This is a read only screen, when accessed by pressing the ENTER key, the user can
view all of the setup information retained in the INFO-CON connector. This informa-
tion tells the user the conversion type, serial number, calibration date, all of the coef-
ficients used for this probe, reference junction type (internal or external) if the probe is
a thermocouple, and the Min and Max temperatures allowed for the probe.
4.5.1.3

Base X

This is the reference value used by the DeltaX function in the auxiliary display. Del-
taX calculates the difference between the current reading and Base X.
4.5.1.4

Aux Displ

This is the Auxiliary display selection. In addition to temperature on the main numeric
output, the user may choose to display an auxiliary measurement for channel T1. For
a resistance thermometer, the user may select to display Ohms, DeltaX, (T1-T2), or
None.
If this display feature is for a thermocouple, the user may choose to display RJ Tem-
perature, mV, (T1-T2), DeltaX or None.
4.5.1.4.1
Ohms
This mode is identified by "W" along the top of the screen. In this mode, the resistance
in Ohms for the temperature measurement appears on the top of the display. This op-
tion is available when the selected channel is reading PRTs or thermistors.