Alpine Opera Eco Installation Manual page 35

Profiling bed
Table of Contents

Advertisement

Item Instruction for testing
18.
Height of side guards above the mattress
at least 220 mm?
19.
Have castors including locking brake
been tested for safe functioning?
20.
Mains cable, connecting cables and
plugs without damage?
21.
Fixture available for safe transportation
of mains plug?
22.
Strain relief of the mains cable and
handset securely attached?
23.
Are all plug-in connections securely
attached? (Washers without damage?)
24.
Are cables laid correctly and safely? (No
damage)
25.
Motor housing and SMPS housing,
mains plug housing without damage?
26.
Are the thrust pipes of the height
adjustment motors undamaged?
27.
Functional test of the handset: can the
buttons be operated properly?
28.
Functional test of handset locking device:
On/Off working correctly?
29.
Testing of initial fault safety by means of
integrated blocking box in handset
30.
9V block battery OK / expiry date
sufficient until next test?
31.
Is the safe working load adhered to?
Comments: .............................................................................................................
Place / Date: ..............................................
Inspected by: .................................................
Next inspection
..............................................
Signature: ..........................................
Eco/Classic Installation Guide and Technical Specification
Comment
Yes No