ACCELLIS C70
TRANSFORMER SPECIFICATIONS
Primary DCR
Secondary DCR
Input Voltage
Input Current
Anode Voltage
WARNING!!
Use extreme caution when taking any cur-
rent readings. Failure to do so can result
in death or serious injury.
FILAMENT TRANSFORMER
BACK VIEW
(1--- 3)
240VAC
COMMON
27.8 OHM
3
2
(1--- 2)
200/208VAC
PRIMARY
FRONT VIEW
23.2 OHM
240V
200/208V
208--- 240
VAC INPUT
Anode Transformers
0.58 Ohm
(5---6) 46.39 Ohm
(5---7) 54.00 Ohm
208---240 VAC, 50/60 Hz
9.0 Amps RMS
4500 VDC Peak @ 0.415 Amps
(4--- 5) 0.20 OHM
4.6VAC SECONDARY
4
5
PRIMARY
1
COMMON
T2 FILAMENT TRANSFORMER
3
4
2
1
5
T3 OR T4
T1 ANODE TRANSFORMER
3
NOT USED
4
1.05uF (60Hz)
2
1.2uF (50hZ)
50Hz
7
60Hz
1
HV
6
CAPACITOR
T1 OR T2
5
15KV
HV DIODE
FIGURE 4
Filament Transformers
208/240 ¦ 10% 50/60 Hz
0.42 Amps @ 208 VAC
0.35 Amps @ 240 VAC
WARNING!!
When replacing the anode transformer,
remove the varnish around the mounting
holes to insure proper grounding.
ANODE TRANSFORMER
REAR VIEW
7
50Hz
60Hz
6
5
HV ANODE WINDING (SECONDARY)
(5--- 6) 46.39 OHM
(5--- 7) 54.00 OHM
FA
F
MAG1
4500 VPK
415 MA
6---5
(1---2) 23.2 Ohm
(1---3) 27.8 Ohm
(4---5) .020 Ohm
4.6 VAC @ 14 Amps
FILAMENT
4
WINDING
3
(SECONDARY)
(1--- 2)
2
208--- 240VAC
(PRIMARY)
1
0.58 OHM