34925T-002 Terminal Block For One-Wire Mode - Agilent Technologies 34921A User Manual

Low frequency multiplexer modules
Table of Contents

Advertisement

34925T-002 Terminal Block for One-Wire Mode

This terminal block with screw- type connections is labeled with the model
number and the abbreviated module name. In addition, space is available
on the label for you to write the slot number.
The 34980A Product Reference CD (shipped with the instrument) contains
a 34925T (1- wire mode) Wiring Log for you to document your wiring
configuration for this module. You can open the wiring log file in
Microsoft® Excel® or Adobe® Acrobat® format.
Agilent 34921A-34925A Low Frequency Multiplexer Modules User's Guide
All modules that connect to the internal DMM are interlock
N O T E
protected. This means that when an installed module is exposed
(no terminal block or cable is connected), the Analog Bus relays are
open and disconnected from the Analog Buses. See
Interlock"
If you are using an Agilent terminal block to connect your DUT to
N O T E
this module be sure to use the 34925T-002 terminal block that
corresponds to the 1-wire configuration mode. An error will not be
generated if you have installed a terminal block that doesn't match
the present module configuration.
34925A 40/80-Channel Optically-Isolated FET Multiplexer
on page 3 for further information.
"Safety
41