Network Analyzer System Operation; Functional Groups Of The Network Analyzer - Agilent Technologies N5230A Service Manual

2-port pna-l microwave network analyzers
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
N5230A

Network Analyzer System Operation

The PNA-L network analyzer generates a phase-locked incident signal and an LO signal
from the internal synthesized source. By means of signal separation, the incident signal is
divided into a reference signal and a test signal.
The reference signal is applied to the receiver group, while the test signal is applied to the
device under test (DUT) and then to the receiver group. The LO signal is applied directly to
the receiver group where it is mixed with the test and reference signals to produce IF
signals for each of the four channels. These IF signals are downconverted and then
sampled and digitally processed.
analyzer system.
Figure 5-1 System Simplified Block Diagram

Functional Groups of the Network Analyzer

The operation of the network analyzer can be separated into major functional groups. Each
group consists of assemblies that perform a distinct function in the instrument. Some of
the assemblies are related to more than one group, and all of the groups, to some extent,
are interrelated and affect each other's performance. The major functional groups are:
Synthesized Source Group
Signal Separation Group
Receiver Group
Digital Processor and Digital Control Group
Power Supply Group
Service Guide N5230-90014
Figure 5-1
is a simplified block diagram of the network
Theory of Operation
Network Analyzer System Operation