Zero Calibration
a.
Turn the indicator on by pressing the C
key.
b.
Set the indicator to the 0.1 range.
c.
Connect the ice bath probe to the indicator
Input jack.
d.
Place the probe in a stable ice bath (0.00C
0.05 C and adjust the ZERO pot (Rb)
located beneath the plug on the rear case
for an indicator display of 0.0 C.
e.
Replace the rubber plug.
Calibration is complete.