Omega Engineering OM5-C User Manual page 20

Isolated signal conditioners
Table of Contents

Advertisement

OM5-BP-16-MUX-C 16 POSITION ANALOG I/O BACKPANEL, MULTIPLEXED
DESCRIPTION
The OM5-BP-16-MUX-C 16 channel backpanel (Figure 5) can accept any
of the OM5 analog modules in any mixture. It can be mounted on the
OMX-1363-C 19-inch metal rack. The OM5-BP-16-MUX-C has two
analog buses; one for analog input and one for analog output. This two-
bus configuration takes advantage of the switch controlled outputs on the
input modules and the track-and-hold inputs on the output modules. A
temperature sensor is mounted on each channel to provide cold junction
compensation for thermocouple input modules (See Figure 6 for
schematic). Field connections are terminated with four screw terminals at
each module site. Up to four OM5-BP-16-MUX-C backpanels may be
daisy-chained. Use OMX-CAB-01-C cable for daisy chaining and OMX-
1315-XX-C cable for connecting to host computer.
26 Conductor Male
Pin Connector P1
(1)
3.47"
P1
(88.1mm)
E1
E2
4
3.05"
3
2
1
(77.5mm)
0
Pin 2
+5VDC Supply
Pin 1
Read Address
Channel
Selection Jumper
Swaged Standoff
Thru Hole 0.15" (3.8mm) Dia.
NOTES:
1. Connector is Berg 65863-075 or equivalent.
Mating Connector is Amp 49958-6 or equivalent.
FIGURE 5. OM5-BP-16-MUX-C Analog I/O Backpanel.
17.4" (442.0mm)
Write Address
Sockets for Current Conversion Resistors, OMX-1362-C
Selection Jumper
1
2
3
4
5
6
7
38
SPECIFICATIONS
Operating Temperature:
–40°C to +85°C
95% relative humidity, non-condensing
Interface Connector:
Field
High Density Screw Clamp, 14 AWG Max
Logic
26-pin, male header connector
Address Input Logic Levels:
Max Logic "0"
0.8V
Min Logic "1"
2.0V
I
Input Current, "0" or "1"
I
RD EN\ or WR EN\ Signal Delay
from Connector P1 to Channels 1-16
Standalone (address 0-15)
Expanded (address 16-63)
17.1" (434.3mm)
11.3" (287.0mm)
7.6" (193.0mm)
3.9" (99.1mm)
8
9
10
11
12
13
14
0.14" (3.6mm)
inside diameter
0.1µA max at 25°C
1.0µA max –25°C to +85°C
51ns at 25°C
64ns at –25°C to +85°C
100ns at 25°C
126ns at –25°C to +85°C
0.15"
(3.8mm)
1.275"
(32.4mm)
10.32 X 0.40
Ground Lug, Including
Nut and Lockwasher
OMX-CJC-C
Cold Junction
Compensation Circuit
0.08"
15
16
(2.0mm)
2.27"
(57.7mm)
0.093"
(2.4mm)
0.75" (19.1mm)

Advertisement

Table of Contents
loading

Table of Contents