Control solution storage conditions
Electromagnetic Compatibility
This meter meets the electromagnetic immunity requirements as per ISO 15197 Annex A. The chosen
basis for electrostatic discharge immunity testing was basic standard IEC 61000-4-2.
In addition it meets the electromagnetic emissions requirements as per EN 61326. Its electromagnetic
emission is thus low. Interference from other electrically driven equipment is not to be anticipated.
Performance Analysis
The performance data for the Accu-Chek Performa system (Accu-Chek Performa meter with Accu-Chek
Performa test strips) were obtained using capillary blood from diabetic patients (method comparison,
accuracy), venous blood (repeatability) and control solution (reproducibility). The system is calibrated
with venous blood containing various levels of glucose. The reference values are obtained using the
hexokinase method. For method comparison, the results were compared with results obtained using
the hexokinase method with deproteinization (automatic analyzer). The hexokinase method is traceable
to an NIST standard.
The Accu-Chek Performa system meets the ISO 15197 requirements.
Weight
Construction
Protection class
Meter type
Approx. 62 g (with battery)
Hand-held
III
The Accu-Chek Performa Meter is suitable
for continuous operation
2ºC to 32ºC
65