Operating humidity
(IEC 60068-2-56).............................. 10 to 90% RH,
Storage humidity
(IEC 60068-2-56).............................. 5 to 95% RH,
Maximum altitude............................. 2,000 m
Pollution Degree (IEC 60664) .......... 2
Shock and Vibration
To meet these specifications, you must panel mount the system and
either affix ferrules to the ends of the terminal wires or use the
NI 9939 backshell kit to protect the connections.
Operating vibration
Random (IEC 60068-2-64)......... 5 g
Sinusoidal (IEC 60068-2-6) ....... 5 g, 10 to 500 Hz
Operating shock
(IEC 60068-2-27).............................. 30 g, 11 ms half sine,
NI 9217 Operating Instructions and Specifications
noncondensing
noncondensing
, 10 to 500 Hz
rms
50 g, 3 ms half sine,
18 shocks at 6 orientations
22
ni.com