Installation - Panasonic MDF-DC700VX Series Operating Instructions Manual

Ultra-low temperature freezer
Table of Contents

Advertisement

■ A location without corrosive substance
Never install the unit in a location where corrosive substance such as sulfur-contained compound may
generate (ex. near a sink). The corrosion of copper pipe of the cooling circuit may result in the failure of
the unit.
■ A location without the possibility of anything fall
Avoid installing the unit in the location where anything can fall down onto the unit. This may cause the
breakdown or failure of the unit.

Installation

1. Remove the packaging materials and tapes
Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the
outside panels are dirty, clean them with a diluted neutral dishwashing detergent (Undiluted detergent can
damage the plastic components. For the dilution, refer to the instruction of the detergent). After the
cleaning with the diluted detergent, always wipe it off with a wet cloth. Then wipe off the panels with a dry
cloth.
Note: Remove the cable tie banding the power supply cord. Prolonged banding may cause the corrosion
of the cord coating.
2. Adjust the leveling feet
It is possible to adjust the length of the leveling feet
by turning them (Fig.1). Lengthen the leveling feet
until the casters are lifted from the floor, and let 3
leveling feet support the front side of this unit. After
that, adjust 3 leveling feet to steady this unit.
3. Fix the unit
Two fixtures are attached to the rear of the frame.
Fix the frame to the wall with these fixtures and rope
or chain.
4. Installation branch circuit breaker
This unit is to be connected to a dedicated circuit protected by branch circuit breaker.
Contact our sales representative or agent.
WARNING
Use a power supply outlet with ground (earth) to prevent electric shock. If the power supply outlet is
not grounded, it is necessary to install a ground by qualified engineers.
Never ground the unit through a gas pipe, water main, telephone line or lightning rod. Such
grounding may cause electric shock in the case of an incomplete circuit.
Shrink
Lengthen
15
Leveling feet
Fig. 1