9. Cleaning and storing the unit

• Clean your blood pressure device carefully with a slightly
damp cloth only.
• Do not use any cleaning agents or solvents.
• Under no circumstances should you hold the unit under
water, as this can cause liquid to enter and damage the
unit.
• If you store the unit, no heavy objects should be placed
on top of it. Remove the batteries. The cuff line should not
be bent sharply.

10. Technical specifications

Model no.
SBM 42
Measurement
Oscillometric, non-invasive blood
method
pressure measurement on the upper
arm
Measurement range
Cuff pressure 0-300 mmHg,
systolic 60-260 mmHg,
diastolic 40-199 mmHg,
Pulse 40-180 beats/minute
Display accuracy
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement inac-
curacy
Memory
Dimensions
Weight
Cuff size
Permissible operating
conditions
Permissible storage
conditions
Power supply
Battery life
Accessories
Classification
11
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
4 x 30 memory spaces
L 134 mm x W 102 mm x H 52,5 mm
Approximately 280g (without batter-
ies and cuff)
22 to 36 cm
+10 °C to +40 °C, ≤ 85 % relative air
humidity (non-condensing)
-20 °C to +50 °C, ≤ 85 % relative air
humidity, 800-1050 hPa ambient
pressure
4 x 1,5 V
AA batteries
For approx. 500 measurements,
depending on the blood pressure
level and/or pump pressure
Instruction for use, 4 x 1.5 V AA bat-
teries, storage pouch
Internal supply, IPX0, no AP or
APG, continuous operation, type BF