Protection class
Control solution storage conditions
Electromagnetic Compatibility
T his meter meets the electromagnetic immunity requirements as per EN ISO 15197 Annex A.
The chosen basis for electrostatic discharge immunity testing was basic standard
IEC 61000-4-2.
In addition, it meets the electromagnetic emissions requirements as per EN 61326. Its
electromagnetic emission is thus low. Interference from other electrically driven equipment
is not to be anticipated.
68
Display
LCD
Dimensions
93 x 52 x 22 mm (LWH)
Weight
Approx. 62 g (with battery)
Construction
Hand-held
III
Meter type
The Accu-Chek Performa Meter is suitable
for continuous operation
2 °C to 32 °C