Error Codes And Symbols - Bayer HealthCare Contour Next EZ User Manual

Hide thumbs Also See for Contour Next EZ:
Table of Contents

Advertisement

®
40
Press ▲ again to view the
30 day average of results
marked with a pre-meal
marker.
Press ▲ again to view the
30 day average of results
marked with a post-meal
marker.
Press ▲ again and you will
see
.
Press M to turn the meter
off, or it will automatically
turn off after 3 minutes.

Error Codes and Symbols

If you do not resolve the problem, contact
Bayer Diabetes Care Helpline at 1-800-268-7200.
What
What It
You See
Means
(will
remain on
Low batteries.
the display)
(will
flash for
10 seconds
Dead batteries. Replace the batteries.
and then the
meter will
turn off)
The meter is
ready to test.
Not enough
blood sample.
E1
Temperature
out of range.
The test strip
has not filled
E2
enough for an
accurate test.
Insufficient
blood drop.
Need Help? 1-800-268-7200
www.diabetes.bayer.ca
What You Should Do
Replace the batteries as
soon as possible.
Proceed with the testing of
your blood.
Add more blood within
20 seconds.
Move to an area that is
within the operating range
for the meter: 5°C – 45°C.
Let the meter adjust to this
temperature for 20 minutes
before performing a test.
Remove the test strip and
retest using a new test
strip. Wait until you see the
flashing blood drop on the
display before testing.
41