Before Calibration Adjustment

Before the Product calibration can be adjusted, you must put the Product into its
Calibration mode menu and enter your password.
To get to Calibration mode:
1. With the Product off, push and hold .
2. Turn the knob to . CAL is shown. The Product is now in Calibration mode.

Enter the Password

To enter the password:
1. Push . The Calibration counter is shown, for example: N002.
2. Push . The display shows "????". The factory default password that must be
3. The top row of buttons corresponds to each number:
4. Push  to go to the first calibration point C-01.

Change the Password

If you change the password and then lose it, see the "Restore the Default
Password" section.
To change the password:
1. Push . The Calibration counter is shown, for example: N002.
2. Push . The display shows "????".
3. Enter the correct password digits.
4. Push . "----" is shown.
5. Enter the new password digits.
6. Push  to enter and store the new password.
14
entered is "1234".
Push  to enter a 1.
Push to enter a 2.
Push  to enter a 3.
Push  to enter a 4.
Push  to enter a 5.
As each number is entered, the cursor automatically moves to the next number.
To enter "1234", push , , , and .
If the correct password is entered, "C-01" is shown. If the incorrect password is
entered, "????" is shown and the password must be entered correctly to go to the first
calibration point, "C-01".