Humidifier Performance; Air Tubing; Pneumatic Flow Path - ResMed AirCurve 10 User Manual

Hide thumbs Also See for AirCurve 10:
Table of Contents

Advertisement

 
Wireless module
Technology used:
It is recommended that the device is a minimum distance of 2 cm from the body during operation. Not applicable to masks,
tubes or accessories
Operating pressure range
S, VAuto:
CPAP:
Supplemental oxygen
Maximum flow:

Pneumatic flow path

Design life
Device, power supply unit:
Cleanable humidifier:

Air tubing:

General
The patient is an intended operator.

Humidifier performance

The following settings have been tested at 22°C ambient temperature:
Mask Pressure
cm H
O (hPa)
2
3
4
10
20
25
1
AH - Absolute Humidity in mg/L
2
BTPS - Body Temperature Pressure Saturated
Air tubing
Air tubing
Material
ClimateLineAir
Flexible plastic and electrical components
SlimLine
Flexible plastic
Standard
Flexible plastic
Heated air tubing temperature cut-out:  41°C
Notes:
 The manufacturer reserves the right to change these specifications without notice.
 The electrical connector end of the heated air tubing is only compatible with the air outlet at the device end and should
not be fitted to the mask.
 Do not use electrically conductive or antistatic air tubing.
 The temperature and relative humidity settings displayed are not measured values.
RH output %
Setting 4
85
85
85
85
85
2G GSM
2 to 25 cm H
4 to 20 cm H
15 L/min (CPAP, S); 4 L/min (VAuto)
1. Flow sensor
2. Blower
3. Pressure sensor
4. Mask
5. Air tubing
6. Humidifier
7. Device
8. Inlet filter
5 years
2.5 years
6 months
Setting 8
100
100
100
90
90
O (2 to 25 hPa)
2
O (4 to 20 hPa)
2
Nominal system output AH
Setting 4
6
6
6
6
6
Length
2 m
1.8 m
2 m
1
2
, BTPS
Setting 8
>10
>10
>10
>10
>10
Inner diameter
15 mm
15 mm
19 mm
English
17