8-4-4 Backing up the Patient Archive and System Configurations
CAUTION
NOTE:
8-4-5 Restoring up the Patient Archive and System Configurations
NOTE:
8-4-6 Recording important settings and parameters
8-4-6-1 Overview
CAUTION
Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3
An error, or a power loss may occur.
Always backup the Patient Archive and the Presets (System
Configurations) before loading the software!
In order to complete a successful restore of the Patient
Database, as needed after a hard disk replacement, or if all the
content on the hard disk has been erased, the images must be
moved away from Vivid T9/Vivid T8 before doing backup of the
Patient Database.
Depending on the location set-up, either move the images to a
remote server or to removable media like DVD or CD discs.
•
Backup the Patient Archive and System Configurations.
For instructions, please see "Data Backup and Restore" in
the User Manual/User Guide.
The system supports back up the presets remotely.
•
Restore the Patient Archive and System Configurations.
For instructions, please see "Data Backup and Restore" in
the User Manual/User Guide.
The system supports restore the presets remotely.
An error, or a power loss may occur.
It is considered to be best practice to always keep a record on
paper of the settings for the Vivid T9/Vivid T8. Verify if it is
current before you start to load software!
Always keep a record of the settings for the Vivid T9/Vivid T8 on
paper. Verify if it is current before starting a software loading! If
needed, record the settings.
This subsection includes descriptions for recording data from
the following screens:
Loading the software