3-9-4 Setup connection to a DICOM server

3-9-4-1 DICOM Server IP address on the ultrasound system
Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3
Vivid T9/Vivid T8 is configured to work with DICOM servers in a
network environment. Images are first saved on the local image
butter on the system. At the end of the examination the images
are sent to the DICOM server via a DICOM spooler and to the
local database, depending on dataflows.
To connect to the DICOM server, the following information has to
be entered in the system.
The DICOM server IP address
The DICOM server port number
The DICOM server AE title (the server application's name)
1. Press Config (F2) and log on as ADM.
2. Select Connectivity and then select Dataflow tab.
Figure 3-31. Select Dataflow
3. Open the Dataflow name pull-down menu.
4. Select the dataflow you want to configure.
In this example, Worklist/Local Archive - DICOM Server/
Int.HD is selected.
Connectivity setup
3-47