Wash Pressure Measurement; Test Point For Measuring Wash Pressure - Miele PG 8581 Operating Instructions Manual

Dental washer-disinfector
Table of Contents

Advertisement

Application technology
Test point for
measuring wash
pressure
36

Wash pressure measurement

The wash pressure can be measured on all mobile units and baskets
with spray arms, injector bars, or other wash connections, e.g., dur-
ing performance tests and validations in accordance with
EN ISO 15883.
On mobile units and baskets with spray arms and additional injector
bars or other wash connections, there is a connection on the injector
bar or a wash connection for wash pressure measurement. The exact
location is described in the respective operating instructions for the
mobile units and baskets.
On mobile units and baskets with spray arms and no additional wash
connections, access for the wash pressure measurement is provided
on the side of the water supply pipe.
Under no circumstances may wash items, irrigation connectors,
etc., be connected to the test point. After the measurement, the
test point must be closed again with the blind stopper.
To measure the wash pressure, replace the blind stopper with a
Luer Lock adapter, e.g., E 447.