Care And Cleaning; Electromagnetic Compatibility (Emc) - Contour Core Sculpting System II Use And Care Manual

Table of Contents

Advertisement

CARE AND CLEANING

To clean the Muscle Stimulator, use a soft cloth lightly dampened with an
alcohol-based cleaning product that does not contain solvents. Solvents
could damage the plastic parts, especially the display.
The Muscle Stimulator is designed to be maintenance free.
There are no user serviceable parts inside.
The Ab Belt is machine washable in warm water/gentle cycle.
You must always follow the instructions on the label when washing the belt
• Do not use any bleach when washing your belt
• Do not dry clean your belt
• Do not tumble dry your belt. Always dry the belt on a flat surface
• Do not dry over anything hot as the belt contains plastic parts.
Ensure the belt is completely dry before use
• Do not iron
IMPORTANT: NEVER REPLACE BATTERIES WHILE THE DEVICE IS ON
Warnings concerning battery handling
Always read and follow the specific instructions of the battery manufacturer.
Note:
• When inserting batteries ensure that battery polarity is correct.
• Do not expose batteries to temperatures exceeding manufacturer's
specifications.
• Operating Temperature Range for Batteries: -20°C to 54°C (-4°F to 130°F)
• Do not store and/or ship this unit with batteries inserted.
• Do not use lithium batteries.
• Do not attempt to recharge alkaline batteries.
• Do not dispose of any batteries in a fire.
• Dispose of batteries in accordance with local regulations.
Please contact your local authorities if you need more information.
• Note that batteries may present burn or fire hazard if short-circuited.
• Improper battery handling may result in explosion, leakage or flames.
• Be sure to remove the batteries when the unit is not used on a
regular basis.
Electrode disposal
Never dispose Gel Pads (new or used) in a fire. Dispose of in accordance
with local regulations. Please contact your local authorities if you need more
information.
Electronic device disposal
Please apply regulations enforced in your country when disposing your
device at its end of life. Therefore, please discard it in a location planned to
this effect, ensuring it is disposed in an environment friendly manner.
28
ELECTROMAGNETIC COMPATIBILITY WARNINGS:
Changes or modifications to this system not expressly approved by the manufacturer may
result in increased emissions or decreased immunity performance of the equipment or system
and could cause EMC issues with this or other equipment. This system is designed and tested
to comply with applicable regulations regarding EMC and shall be installed and put into
service according to the EMC information stated as follows. Use of portable phones or other
radio frequency (RF) emitting equipment near the system may cause unexpected or adverse
operation. The equipment or system shall not be used adjacent to, or stacked with, other
equipment. If adjacent or stacked use is necessary, the equipment or system shall be tested
to verify normal operation in the configuration in which it is being used.
Guidance and manufacturer's declaration – electromagnetic emissions
The EQUIPMENT is intended for use in the electromagnetic environment specified below. The
customer or the user of the EQUIPMENT should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
RF emissions
Group 1
The EQUIPMENT uses RF energy only for its internal
CISPR 11
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
Class B
The EQUIPMENT is suitable for use in all establishments
CISPR 11
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Guidance and manufacturer's declaration – electromagnetic immunity
The EQUIPMENT is intended for use in the electromagnetic environment specified below. The
customer or the user of the EQUIPMENT should assure that it is used in such an environment.
Immunity tests
IEC 60601
Compliance level
Test level
Electrostatic
± 6 kV
± 6 kV Contact
discharge (ESD)
Contact
discharge ± 8 kV
IEC 61000-4-2
discharge
Air discharge
± 8 kV Air
discharge
Electromagnetic environment -
guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.