MRC L3201LED User Manual

Led epifluorescence microscope

Advertisement

1) While the push rod to transmitted or reflected is to the working
position "R", fluorescent microscopy can be used.
2) Reduce the field diaphragm, If it is not centered, move it to the
center using the special tool. Adjustment can be done by opening the
field diaphragm.
3) The push rod to direct image to binocular eyepieces or camera port
to the upper left of the trinocular head will send 100% of light to the
binocular eyepieces or to the top port. It satisfies high-grape
photography, see figure 7.

Preventative Maintenance

1. Cleaning frame and stage
Disconnect the plug from mains socket before cleaning. Clean the frame
and stage with a soft cloth moistened with a mild detergent solution. Be
sure the instrument is dry before using.
2. Cleaning optical parts
Microscope eyepieces and objectives are coated. They should not be
wiped while dry as dirt or dust may scratch the coating. It is best to
remove parts from the frame prior to cleaning. Always blow loose dust
away first. Use cotton swabs or lens tissue moistened with a lens cleaner
or a small amount of alcohol, then wipe the surface clean with a good
quality lens tissue. Solvents such as Xylene should NOT be used as
cleaner.
3. Cleaning 100X oil immersion lens
The immersion oil should be removed from the lens at the end of each
workday using cotton swabs or lens tissue moistened with a lens cleaner
or a small amount of alcohol.
11
LED Epifluorescence Microscope
L3201LED
User Guide
HAGAVISH 3,HOLON, ISRAEL
TEL: 972-3-5595252, FAX. 972-3-5594529
mrc@mrclab.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the L3201LED and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for MRC L3201LED

  • Page 1 3) The push rod to direct image to binocular eyepieces or camera port to the upper left of the trinocular head will send 100% of light to the L3201LED binocular eyepieces or to the top port. It satisfies high-grape photography, see figure 7.
  • Page 2: Figure

    swing the attachment lens to optical center. Congratulations and Thank You for your purchase of Condenser Alignment – to center the Abbe condenser into the optical light path, see figure 6: The Microscope 1) Close the field iris diaphragm. 2) Observe the image using the 5X This precision instrument has been designed to require a minimum of optical objective and you will see an and mechanical maintenance.
  • Page 3: Table Of Contents

    1) Diopter Adjustment Proper correction for individual vision is Table of Contents accomplished via the diopter adjustment located at the left eyepiece. Using the 10X objective, bring an Diopter Adjustment Ring image into focus with your right eye Getting Started ....................3 only.
  • Page 4: Getting Started

    3) Reflection Illumination Press the transmitted / reflection illumination Getting Started This reference guide is written base on whole series, it covers some capable toggle switch on the side of “ Ⅱ ”, which means the reflection optional attachments and functions, you may read the similar operation what illumination is used.
  • Page 5: Components

    Adapter: Input: 110V~240V 50/60Hz Components Output: 12V Setup Instructions Remove all parts from their packing materials and retain the packaging in the event you need to transport the product. Connect the power cord to power outlet. Basic Operation Illumination controls, see figure 3 1) The external power switch and transmitted reflection...
  • Page 6: Technical Specifications

    Technical Specifications Inclined 30°. Trinocular: Objectives: Plan achromatic objectives Magnification N.A. W.D. (mm) 0.10 19.75 0.25 0.65(Spring) 0.66 100X 1.25(Spring, Oil) 0.36 Quadruple/Quintuple nosepiece ball-bearing reversed nosepiece with positiv Nosepiece: 10X Plan wide field, focal length 25mm, Field Φ20mm. Eyepieces: 210mm X 140mm, range of motion 75mm X 50mm.

Table of Contents