Andon Blood Glucose Monitoring System Operation Manual
DANGER-Misuse of electrical equipment can cause electrocution, burns, fire
Basic safely precautions should always be taken, including all those listed below.
Close supervision is necessary when the Blood Glucose Monitoring System is used by, on, or
near children or invalids.
READ THIS BEFORE USING THE EQUIPMENT
 Check that the voltage setting matches the supply voltage.
 Switch off the MAINS supply switch immediately after use.
 Do not placement the Blood Glucose Monitoring System in liquid, nor put it where it
could fall into liquid.
 Use the Blood Glucose Monitoring System only for the purpose described in the
instructions for use.
 Do not use accessories which are not supplied or recommended by the Andon.
 Do not use the Blood Glucose Monitoring System if it is not working properly, or if it has
suffered any damage.
 Do not let the Blood Glucose Monitoring System come into contact with surfaces which
are too hot to touch.
 Do not block test ports nor place the Blood Glucose Monitoring System on a soft surface
which might block them, and keep test port free from lint, hair, fluff .etc
 Do not place anything on top of the Blood Glucose Monitoring System.
 Unless specifically instructed to do so by the instructions for use, and do not drop or put
anything into any opening in the Blood Glucose Monitoring System.
 Do not use the Blood Glucose Monitoring System where aerosol sprays are being used,
or where oxygen is being administered.
 The machine clean and taking good care of at the regular intervals , eliminating a
bloodstain , spreading crumbs , dust and so on.
 The blood glucose monitoring system has gone through the EMC test.
 The device can be reused, do not discard anywhere.
 The system which belongs to List B has an internal power.
 Do not use the device in the condition where mix of flammability air.
Model: AG-605A
IMPORTANT SAFETY INSTRUCTIONS
and other HAZARDS
KEEP THESE INSTRUCTIONS
1

Advertisement

loading
Need help?

Need help?

Do you have a question about the AG-605A and is the answer not in the manual?

Questions and answers

Summary of Contents for Andon AG-605A

  • Page 1  Use the Blood Glucose Monitoring System only for the purpose described in the instructions for use.  Do not use accessories which are not supplied or recommended by the Andon.  Do not use the Blood Glucose Monitoring System if it is not working properly, or if it has suffered any damage.
  • Page 2: Intended Use

    Product Brief Instruction AG-605A blood glucose Monitoring System is designed to provide an easy, accurate method for determination of your blood glucose level within 5 seconds. 350 test records with respective date and time help to review the blood glucose trend easily.
  • Page 3 The blood monitoring system is not intended for use on artery blood, neonates’ serum and plasm. Caring for your AG-605A Blood Glucose Monitoring System Clean the meter Please store the meter in the caring case after each use. A cloth dampened with water and mild detergent can be used to wipe down the outside of the meter.
  • Page 4 Note: Discard the used battery according to the local law. Test Strip The Test Strips are designed to work only in conjunction with the Glucose Meter. Each strip can be used only once. The Glucose AG-605A Test Strip consists of the following parts: Test Strip Vial label...
  • Page 5 Note: Please check test strip indication for use, which indicates test strip’s use life, storage condition and operational step. Setting Up the Date, Time and Unit of Measure 1. Start with the Meter off. Long press the S button. The Meter is now in the setting mode.
  • Page 6  Before code the meter, make sure the test strip have not expired. Testing Your Blood Glucose Level Before testing, you will need the following items:  Glucose AG-605A Meter  AGS-1000 test strip  Adjustable Automatic Lancing Device  Sterile Lancet...
  • Page 7 Step 5 When the blood drop symbol flashes on the meter, it means that you can start testing. You have 3 minutes to apply the blood sample. If you do not apply the sample within 3 minutes, the meter will turn off automatically. Obtain a Blood Sample Sample may be obtained from finger as the following procedure: Remove the end-cap of the lancing device by gently turning and pulling...
  • Page 8 Test Strip is completely filled with your blood sample. Step 9 The Glucose AG-605A results range between 20 mg/dL to 600 mg/dL(1.1 mmol/L to 33.3 mmol/L). If a “HI” message appears on your display, your Glucose Meter has detected that your blood glucose level is higher than 600mg/dL or 33.3mmol/L.
  • Page 9 meter will turn off automatically or you can long press the S button after remove the used test strip to turn the meter off immediately or insert another test strip to begin a new test. NOTICE: To remove the Test Strip from the Glucose Meter for proper disposal , We recommend using a tissue paper to remove the Test Strip form the Glucose Meter.
  • Page 10  Handle the Meter with care. Dropping or throwing the Meter may cause damage to the device.  Don't expose the andon Blood Glucose Monitoring System, Test strip, Control solution, to extreme conditions, such as high humidity, heat, freezing cold or dust.
  • Page 11 Display Messages Display Message What To Do? Meter is ready to accept a Please place a drop of blood sample. blood onto the test well of test strip Meter is ready to measure Please place a drop of under the CTL condition. blood onto the test well of test strip Five-second countdown.
  • Page 12 Display the result Do nothing Demonstrates the memory Do nothing the serial number Memory result Do nothing demonstration No memory in meter Do nothing Result display is end Do nothing...
  • Page 13 Battery power is low Please change all the batteries Meter was not finished Please contact us calibration Use a used test strip Please use a new strip and re-test Meter internal memory unit Reinsert the CODE bottle, error please contact us if it appear for three time.
  • Page 14 Environment temperature Please test at a lower is higher than 40℃(104℉) temperature environment Problem Cause Solution Display remains 1. Batteries may be 1. Install the bank after the Test installed improperly. batteries correctly. strip has been 2. Battery power is too low 2.
  • Page 15 15minutes before performing a new test. The Meter Test Strip has not been Use a new Test Strip countdown did not inserted correctly. and redo the test. start.
  • Page 16 Anti-garbage Lotus Global Co., Ltd 47 Spenlow House Bermondsey London SE16 4SJ Andon Health Co., Ltd. Register Address: No.3 JinPing Street, Ya An Road, Nankai District, Tianjin, China, 300190 Producing Address: No.3 and No.10 JinPing Street, Ya An Road, Nankai...