John Bunn Neb-u-Tyke RR JB0112-064 User Manual

Compressor nebulizer

Advertisement

Quick Links

Neb-u-Tyke RR
COMPRESSOR NEBULIZER
MODEL: JB0112-064
USER'S MANUAL
Read this manual before operating the nebulizer
Save these instructions for future reference
Caution:
Federal law restricts this device to sale by or on order of a physician, or any other
practitioner licensed by the law of the State in which he practices to use or order
the use of this device.
JB0112-064-INS-LABRevB06

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Neb-u-Tyke RR JB0112-064 and is the answer not in the manual?

Questions and answers

Summary of Contents for John Bunn Neb-u-Tyke RR JB0112-064

  • Page 1 Neb-u-Tyke RR COMPRESSOR NEBULIZER MODEL: JB0112-064 USER’S MANUAL Read this manual before operating the nebulizer Save these instructions for future reference Caution: Federal law restricts this device to sale by or on order of a physician, or any other practitioner licensed by the law of the State in which he practices to use or order the use of this device.
  • Page 2: Table Of Contents

    Contents Introduction ............3 Important Safeguards .
  • Page 3: Introduction

    Introduction Congratulations on your purchase of the John Bunn Piston Compressor Nebulizer. Thanks for trusting us. Graham Field guarantees that this product has been manufactured with high quality materials and fulfills all the standing safety standards, both national and international.
  • Page 4: Features

    8. Use the unit on stable surfaces. Never use while sleeping or drowsy. 9. Do not submerge the nebulizer in water. 10. Use the unit intermittently. Maximum operating period: 30 minutes. It must stand unused for 30 minutes between uses. 11.
  • Page 5: Unit Assembly

    Unit Assembly IMPORTANT: Do not pour more than 10 ml.
  • Page 6: Operating Instructions

    2. Remove filter cap by grasping it firmly and pulling out the unit. 3. Change the dirty filter of the cap with small, pointed object. 4. Insert a clean John Bunn filter. Extra filters are available. 5. Push filter cap with new filter into position.
  • Page 7: Cleaning And Maintenance

    Cleaning and Maintenance Important: Before using the nebulizer for the first time or after each use, the accessories must be cleaned as follows: clean all the accessories, except the tubing, with a warm water and mild detergent solution. Rinse with warm tap water for 30 seconds to remove detergent. Soak all the parts in a mixture of three parts water to one part white vinegar for 30 minutes in a container, or use a bacterial germicide disinfectant, which can be obtained in drugstores.
  • Page 8: Technical Specifications

    Technical Specifications Method of nebulization ....... . . Dry piston compressor Operation voltage .
  • Page 9: Trouble Shooting Guide

    Trouble Shooting Guide If something out of the normal happens during operation, please check and correct the following points: Symptom Possible cause Suggested solutions 1-Check the plug is 1-Incorrectly connected to properly connected Although the power source device is on, it 2-Check that power is on 2-No power does not work...
  • Page 10 5 years from the date of purchase. This warranty does not cover failures resulting from misuse, abuse, alteration of the John Bunn Nebulizer System, use of improper components or failure to comply with the operating instructions.
  • Page 11 U.S.A., Corporate Headquarters: Graham-Field Health Products 2935 Northeast Parkway Atlanta, Georgia 30360 telephone: 800-347-5678, 770-447-1609 fax: 800-726-0601, 678-291-3232 http://www.grahamfield.com JB0112-064-INS-LABRevB06 NFIL/31026...