Omega AQM-101 Instruction Sheet page 2

Iaq formaldehyde (hcho) logger

Advertisement

NOTE:
If software message box show as Pic2.Follow the next steps.
"
""
"
1.Click My computer Content in hardware content, click
"
"
Device manager to find the listed (COMx) listed under Ports
(COM & LPT as below with red line. If you may Prolific
USB-to-Serial Comm. Port , it means the logger is connected
to computer properly.
2. Remember the Comm port number and go back to software.
3 Click "Setting" "Comm port " select the port number as you
.
see from step 1, then press OK button.
Logger Setting:
(Pic.5)
(Pic.3)
(Pic.4)
8
Clear Data:
The data is stored in the memory till next setting start, meter
will clean previous data automatically. So if you start next
data logging, there is no way you can find the data you
haven't stored.
(pic.8)
(pic.7)
Line color selection:
1.Click the color beside the reading(Temp. RH. HCHO),
color box appear.
2.Move mouse to review and select color.
3.Click to save, the color appear on line and bar.
Zoom in/out&Move
=
+
Screen graph zoom in
Hold
Click
Ctrl
Mouse
+
=
Hold
Click
Screen graph zoom out
Ctrl
Mouse
+
Hold
=
Click
Move screen graph
Ctrl
Shift
Mouse
Press
or
= Read each measurement .
NOTE: May follow the line color selecting, the graph size will
go back to default.
12
BASIC:(Pic.3)
1:ID setting
(Pic.2)
Maximum 20 characters , name your logger press OK
button to confirm
Please type in capital. Space is unacceptable.
2:Current Date/Time
Clock setting, the system automatically shows current date and
time of your pc. Press OK to confirm.
(Note: Schedule date & time refer to current pc date and time)
Please click OK to sync the meter's date and time
every time.
ALARM&SENSOR:(Pic.4)
1.Appear the parameter.
HCHO model default with Temp. and RH sensor.
SETTING:(Pic.5)
1. Set sample point (K=1,000 multiple basis within total
memory points).
2.Set sample rate, start mode to record:
Select "hour(1~6), "minute(0~59)","second(0~59)" time interval.
Select 5 start modes from:
Immediate (Logger start recording immediately)
Real-time(PC is always connected with the logger)
Key start/off (Start/stop recording by long pressing
SEL/R button)
Roll-over (when memory are full, logger covers the earliest memory
automatically)
Schedule (Select the date and time meter will start recording.)
3.Select Unit :
Select Temperature unit °C/°F for recording,
press OK to confirm.
9
SPECIFICATIONS
AQM-101
0.00-5.00 ppm (mg/m
Meas. Range
Audible alarm
YES (Adjustable)
Interfacce
USB
LED(Green.Red)
YES
Temp. range
0~50°C (32~122°F )
RH%range
10 ~ 90%RH
Resolution
0.01ppm(HCHO) ,1°C /°F , (1%RH)
Accuracy
±1°C/2°F (Temperature), ±5% (10~90%RH)
±5% of reading + 0.03ppm (mg/m3)
( Under vaccum ,ventilated standard gas chamber
). Meet NOISH acceptance criterion within +/-25%
Accuracy(HCHO)
of the true value at th e95% confidence level.
Example: OSHA Formaldehyde regulation
0.10ppm ( EPA 0.08ppm), accuracy within
0.025ppm ( 0.075~0.125ppm)
Response time
T90: Less than 30 seconds
Memory
32,000 points
Sensor life time
Expected 3 years
TROUBLE SHOOTING
(1) When meter appears break character, please find
out if meter ever dropped to the floor. If "yes",
please contact with local distributor for technical service.
(2) Error codes:
Er1 : HCHO/Temp./RH sensor is failed
Er2 : HCHO/Temp./RH Over range
13
Schedule start mode:
Select the date and time meter will start recording.
The recording start time as software's current time not
meter's. Please set up the current time correctly.
Real-time start mode:
1.After set start mode at real-time,datalogger starts login records.
When logger records, the green fan at left corner shows it is
under recording.(Meter should be connected with pc all the time,
)
under this mode.
2.In real-time mode, it shows "STOP" button instead of
"DOWNLOAD" at the right corner. Press STOP to stop
real-time recording.
Before exit software data has to be downloaded.
Key-start/off mode:
Select the mode to record once long press
S
E
/ L
R
button for 3 seconds. Long press
button again to stop recording. Data has to be
downloaded before next key-start. LCD
display appear "Logging" at left corner.
NOTE:
After key off the recording, the data has to download
before restart the Key-Start function.
Roll-over mode:
After Roll-over mode is selected. Press OK, the meter will start to
record. When the data is full, it will cover the earliest memory.
After download the data , long press
without reset up each parameter and selection.
Logger Status:
Before download data, press
LOGGER status
Records, Sample rate, Unit.
NOTE:
Once
Download
button is pressed, recording is
inactivated.
10
WARRANTY
This instrument is warranted for two years from the date of
purchase(One year limited warranty applies to cables). A Return
3
)
Authorization letter must be issued before returning for any reason.
This warranty does not apply to defects resulting from action of
the user such as misuse, abuse, alteration neglect, improper
wiring, improper maintenance or repair, or unauthorized
modification, damage resulting from leaking batteries, operation
outside of specification.
During the warranty period the manufacturer reserved the right to
decide either to repair or replace the product. The meters are to be
returned along with good packing to prevent any damage in
shipment and insured against possible damage or loss.
The two years' warranty doesn't apply to:
Accessories and batteries (not covered by warranty).
Claims is not acceptable for improper use (including
adaptation to particular applications not foreseen in
the instructions manual) or improper combination
with incompatible accessories or equipment, or by
previous attempts for repair carried out by none
skilled or unauthorized personnel.
Sensor damaged by user improper use
(Watering,Burning,Condensation....)
HCHO LEVELS AND GUIDELINES
Non-enforced reference levels:
a) 0.03 ppm-Average outdoor level.
b) 0.10 ppm-Recommend upper limit for residences by
ASHRAE , ANSI,EPA,NIOSH for STEL.
c) 0.40 ppm-Recommended upper limit for manufactured homes.
d) 0.50 ppm-OSHA workplace limit.
e) 0.75 ppm-OSHA TWA limit.
f ) 0.80 ppm-Level at which most people first detect odor.
g) 2.00 ppm-OSHA STEL limit.
14
Logger function:
Download data:
Press DOWNLOAD button, data download in few seconds.
Graph shows automatically
NOTE:
Please do not press DOWNLOAD button if you won't
stop record. You can press LOGGER STATUS button to see
record points.
Retrieve file:
Press RETRIEVE FILE button to select file and show the graph.
Save data:
Press SAVE button to save txt.or csv.file. After save as txt.
may print as pdf. file if needed.
Long press
View Data Table:
S
E
/ L
R
Press View button to get data table with details.(pic.6)
Print graph:
Press PRINT button and print graph. Follow the print setting
below to print full screen graph. Press OK to print pdf file.(pic.7)
Print Table Data:
Press table data, the screen shows the
record range, enter the start number to the
SEL/R
button to record
end number (For example: 1~500 or
200~1000...or leave it blank for printing
all data ).(pic.8)
to show ID,
Save each pdf. files with different names.
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials
and workmanship for a period of 13 months from date of purchase. OMEGA's
WARRANTY adds an additional one (1) month grace period to the normal one (1)
year product warranty to cover handling and shipping time. This ensures that
OMEGA's customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation.
OMEGA's Customer Service Department will issue an Authorized Return
(AR) number immediately upon phone or written request. Upon examination
by OMEGA, if the unit is found to be defective, it will be repaired or replaced
at no charge. OMEGA's WARRANTY does not apply to defects resulting from
any action of the purchaser, including but not limited to mishandling,
improper interfacing, operation outside of design limits, improper repair, or
unauthorized modification. This WARRANTY is VOID if the unit shows evidence of
having been tampered with or shows evidence of having been damaged as a
result of excessive corrosion; or current, heat, moisture or vibration; improper
specification; misapplication; misuse or other operating conditions outside of
OMEGA's control. Components in which wear is not warranted, include but are not
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products.
However, OMEGA neither assumes responsibility for any omissions or errors
nor assumes liability for any damages that result from the use of its products
in accordance with information provided by OMEGA, either verbal or written.
OMEGA warrants only that the parts manufactured by the company will be as
specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR
REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR
IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES
INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR
A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF
LIABILITY: The remedies of purchaser set forth herein are
exclusive, and the total liability of OMEGA with respect to this order,
whether based on contract, warranty, negligence, indemnification,
strict liability or otherwise, shall not exceed the purchase price of the
component upon which liability is based. In no event shall OMEGA be liable
for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it
be used: (1) as a "Basic Component" under 10 CFR 21 (NRC), used in or with any
nuclear installation or activity; or (2) in medical applications or used on humans.
Should any Product(s) be used in or with any nuclear installation or activity,
medical application, used on humans, or misused in any way, OMEGA assumes
no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, and,
additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from
any liability or damage whatsoever arising out of the use of the Product(s) in such
a manner.
(pic.6)
11