Mrc laboratory water purification system

Advertisement

Quick Links

Laboratory Water Purification System
Instruction Manual
Model: Master-S UVF
The instruction manual should be read before installation
Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com
MRC.VER.01-8.13

Advertisement

loading
Need help?

Need help?

Do you have a question about the Master-S UVF and is the answer not in the manual?

Questions and answers

Summary of Contents for MRC Master-S UVF

  • Page 1 Laboratory Water Purification System Instruction Manual Model: Master-S UVF The instruction manual should be read before installation Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC.VER.01-8.13...
  • Page 2: Specification

    RO water and Ultrapure water for science research. So, it will benefit your work. For the water purification system’s maximum efficiency, it is suggested that the user manual be read before installation. Any question in the installation process, please contact our technology engineers or dealers. 2. Specification Master-S UVF Model 18.2 MΩ.cm Resistivity <3ppb...
  • Page 3 Water Purification System Instruction Manual 3. Master-S UVF Water Flow Chart 4. Working Environments  Inlet water: Tap water (TDS<200ppm will be suggested). If inlet water TDS>200ppm, pretreatment is recommended. Water with higher TDS will affect the quality of outlet water and life of purification cartridge.
  • Page 4 Water Purification System Instruction Manual  Specific installation steps are as follows: (1). Connect to tap water-1st step: Shut off the chief switch of tap water, dismantle the tap, connect the water source with 3-direction-links in the accessories, and then connect the tap with 3-direction-links, connect another direction of 3-direction-links with plastic ball valve.
  • Page 5 Water Purification System Instruction Manual Installation Guide Chart RO water outlet: Ultrapure water outlet: Application: Application: ware washing,agricultural GC,HPLC,IC,ICP expe,general biological expe,aquatic PCR,weather analyse products feeding,inlet water for ultra Amino acid analyse pure water machine,inlet water for Reagent preparation sterilizer/ t&h chamber Terminal filter UF membrane drain UF drain valve...
  • Page 6: Usage Guide

    Water Purification System Instruction Manual 6. Usage Guide All data have been set in the factory. The machine will operate smoothly without any data-setting. All work state will display in the LCD. If there is abnormal state, the system will alarm automatically. If data modification is necessary, specific step is in “Microcomputer Controller”.
  • Page 7 Water Purification System Instruction Manual  The usage to keep high quality pure water: (1). The Ultrapure water is easily polluted by surrounding environment. So getting fresh pure water is suggested. (2). Keep source water tank from sunlight for microbe’s reproducing. (3).
  • Page 8 Water Purification System Instruction Manual inlet water is ok, system goes on flushing RO membrane. (5). After “flush” state, system switches to “pure water produce” state (picture 5) to produce pure water. Meanwhile check the state of inlet water and water tank. When tank is full with RO water (picture 6), system will warn.
  • Page 9 Water Purification System Instruction Manual password(In this process, if password is not right, system will warn.) At last press “OK” key-press to switch to menu interface(picture 15), After switching to menu interface, through pressing “MENU” key-press, shift cursor to corresponding item menu, then press “OK” key-press to enter corresponding item. picture 14 picture 15 (1).
  • Page 10 Water Purification System Instruction Manual resistivity < The “Min”, system will warn to replace the Ultrapure polishing resin cartridge. When current test parameter of resistivity < The “Min”, system will stop warning. EC: electrode parameter setting (Parameter setting is OK in factory. Additional parameter setting is unnecessary.) (3).
  • Page 11: Trouble Diagnosis

    Water Purification System Instruction Manual Measure unit: TDS(total dissolved solid, ppm) The third, Resist, monitors Ultrapure water’s quality (Ultrapure water). Measure unit: Resistivity (MΩ.cm) Remarks: Conversion relations between TDS and conductivity rate(µs/cm): rate(µs/cm) ≈ If TDS<50ppm. conductivity TDS×2 rate(µs/cm) ≈ conductivity If TDS>200ppm.
  • Page 12 Water Purification System Instruction Manual Ⅲ. Motor working but no water RO membrane is blocked or pump’s pressure lost. For the former, change the RO membrane; For the later, check whether expiring of pump is caused by long time overload or blocking of the cartridge. Repair motor. Ⅳ.Motor works ever and again The motor works ever and again.