Advertisement

Technical Manual
CardioMessenger
-S TLine
®
Transmitter for BIOTRONIK Home Monitoring
®

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CardioMessenger-S TLine and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for BIOTRONIK CardioMessenger-S TLine

  • Page 1 Technical Manual CardioMessenger -S TLine ® Transmitter for BIOTRONIK Home Monitoring ®...
  • Page 2 Light 1: Call back light (yellow, page 22) Light 2: Operating light (yellow/green, page 21) Slot for the brief instructions guide (page 20) Use only original power supply units (page 16) Connections for the landline (page 8 - 15) Connection for the power supply (page 16)
  • Page 3: Table Of Contents

    Switching off the system....23 Debugging ......24 371017--B CardioMessenger-S TLine...
  • Page 5: Introduction

    Your device is equipped with a special cardioverter-defibrillator (ICD) with the transmitter (1). Usually at night, the additional BIOTRONIK Home Monitoring transmitter sends daily information on  function. With Home Monitoring, the state of your heart to the patient device, your your heart's health and your device are CardioMessenger (2).
  • Page 6 The CardioMessenger collects the information sent by the device and automatically transmits it to the BIOTRONIK Service Center (4) as encoded messages via a landline connection (3). Here, the messages are decoded and made available on a protected web site. Your...
  • Page 7: Check The Completeness Of The Delivery

    Check all components for visible damage before using them. Use only components which are undamaged. The delivery includes the following: CardioMessenger-S TLine Power supply unit with electricity cable and  DC plug Transport packaging Keep the transport packaging and use it if you return the CardioMessenger to your physician.
  • Page 8 Warning!  Use only the original parts included (for details, see "Technical data", page 37). Other equipment may impair proper functioning of the CardioMessenger and increase the emitted interference and the CardioMessenge's susceptibility to interference. The label on the back side of the Technical manual with brief instructions guide CardioMessenger indicates the approved power supply:...
  • Page 9: Setup

    Please take the following into consideration: and high humidity. Do not place it in the bathroom. Place the CardioMessenger in such a way that the lights can easily be viewed during the day. If the lights disturb you in your CardioMessenger-S TLine Setup...
  • Page 10: Connection

    Connection Connection to the landline In both cases, you can use your phone as usual. The CardioMessenger transmits the  Note  Home Monitoring data via landline. The CardioMessenger is designed to There are two ways to connect the operate with standard analog phone lines CardioMessenger to the landline: providing Plain Old Telephone Service (POTS).
  • Page 11 1. On the back side of the CardioMessenger: Remove the sealing plug from phone inlet 1. Phone inlet 1 is labeled with the following symbol: 2. Unplug the telephone cable from the telephone wall jack. CardioMessenger-S TLine Connection...
  • Page 12 3. Connect the free end of the telephone cable 4. Remove the sealing plug from phone inlet 2. to phone inlet 1 of the CardioMessenger. Phone inlet 2 is labeled with the following symbol: The telephone is now connected to the CardioMessenger.
  • Page 13 6. Connect the other end of the supplied connect one end to phone inlet 2. telephone cable to the telephone wall jack. 7. Check the telephone connection. Lift and listen to the receiver. The usual dialing tone should be audible. CardioMessenger-S TLine Connection...
  • Page 14 The telephone and CardioMessenger are now connected to the landline. As a next step, connect the CardioMessenger to the power supply (see "Connection to the power supply", page 16). Connection...
  • Page 15 1. On the back side of the CardioMessenger: Remove the sealing plug from phone inlet 1. Phone inlet 1 is labeled with the following symbol: CardioMessenger-S TLine Connection...
  • Page 16 2. Take the supplied telephone cable and 3. Connect the other end of the supplied connect one end to phone inlet 1. telephone cable to the telephone wall jack. Connection...
  • Page 17 The CardioMessenger is now connected to the landline. As a next step, connect the CardioMessenger to the power supply (see "Connection to the power supply", page 16). CardioMessenger-S TLine Connection...
  • Page 18: Connection To The Power Supply

    Connection to the power supply 1. Connect the DC plug at the end of the electricity cable to the port on the left side Connect the CardioMessenger to the power of the CardioMessenger. supply. The outlet to be used has to be easily accessible.
  • Page 19 2. Connect the power supply unit to the outlet. See that the outlet is not controlled by a light switch. This will prevent you from turning off the CardioMessenger accidentally. CardioMessenger-S TLine Connection...
  • Page 20: Self-Test

    CardioMessenger checks Note  the connection to the It is a sign of malfunction if the operating BIOTRONIK Service light remains illuminated yellow for more Center. Only the than 15 minutes or if it does not light up at operating light all.
  • Page 21 If you are operating CardioMessenger and If you are operating the CardioMessenger on a telephone using one telephone wall jack free telephone wall jack (variant B), the readily (variant A), the readily connected connected CardioMessenger looks as follows: CardioMessenger looks as follows: CardioMessenger-S TLine Connection...
  • Page 22: Insert Brief Instructions Guide

    Insert brief instructions guide 3. Insert the brief instructions guide into the slot on the left side into the housing of the Attached to the back cover of this technical CardioMessenger. manual is a removable instructions guide. This guide helps to inform you quickly about the function and colors of the lights.
  • Page 23: Operate

     The CardioMessenger automatically receives the information from your device and transmits Note  it to the BIOTRONIK Service Center. The functions of your device are not affected Note at any time by the CardioMessenger,  It indicates a malfunction if the operating...
  • Page 24: Call Back Function

    Call back function Perform call back Call your cardiologist or the physician who With the aid of the call back light, your usually carries out the follow-up examinations physician can ask you to call him. He can turn as soon as you realize that the call back light is this light on via the landline.
  • Page 25: Switching Off The System

    The CardioMessenger performs the self-test. Afterwards, the operating light will illuminate green and the call back light will no longer be blinking. Note  For details on the subject of "Connection to the power supply", see page 16. CardioMessenger-S TLine Operate...
  • Page 26: Debugging

    Debugging You can identify issues with your CardioMessenger by use of the lights on the front side. Light Behavior Operating status Call back Normal operation; no malfunction light Yellow illumination Your attending physician is asking for a call (For details on the subject of "Call back function", see page 22.);...
  • Page 27: Cardiomessenger-S Tline

    CardioMessenger checks the connection to the the lamp on. BIOTRONIK Service Center. The operating If you do not find any errors, contact your light switches to green once the connection cardiologist or the physician who usually test is sucessful.
  • Page 28 The connection check can take up to back light is off. CardioMessenger cannot 15 minutes. connect to the BIOTRONIK Service Center. 1. Make sure that the one end of the telephone cable is properly inserted into the telephone wall jack.
  • Page 29: Cardiomessenger-S Tline

    It checks the connection to the providing Plain Old Telephone Service BIOTRONIK Service Center. The operating (POTS). Compatibility with other phone lines light switches to green once the connection or services (e.g., Voice over IP (VoIP) via test is sucessful.
  • Page 30: Cleaning, Maintenance, And Disposal

    Cleaning, maintenance, and disposal Cleaning Maintenance Keep the CardioMessenger clean and away The CardioMessenger is intended for from dirty or dusty environments. continuous, automatic operation. Once correctly installed, ongoing actions by you are Use a soft, lint-free cloth for cleaning. typically not required (e.g., no maintenance is Caution! ...
  • Page 31 If no longer required or if defective, please return the CardioMessenger and all other supplied parts to the physician. Your physician will return all parts to BIOTRONIK. BIOTRONIK ensures the disposal in accordance with the national design of the European guideline 2002/96/EC on electric ...
  • Page 32: Precautionary Measures

    Precautionary measures The CardioMessenger is a medical device and Warning!  therefore complies with strict requirements for Use only the original parts included (for their development, manufacture and testing. details, see "Technical data", page 37). Other equipment may impair proper Please take the following precautions: functioning of the CardioMessenger and Warning! ...
  • Page 33 – Temperatures above 40°C (104°F)  (for example from direct sunlight and strong halogen spotlights) – Temperatures below 10°C (50°F) – Solvents, acids, detergents, and lyes Caution!  The CardioMessenger may only be opened and repaired by authorized trained personnel. CardioMessenger-S TLine Precautionary measures...
  • Page 34 Precautionary measures...
  • Page 35: Guidelines

    Services and must accept any interference transmission from this transceiver will be free received, including interference that may from interference. cause undesired operation. 1) Federal Communications Commission for Medical Implant Communications Service CardioMessenger-S TLine Guidelines...
  • Page 36 The FCC ID number for the CardioMessenger cause harmful interference to radio is QRICM08V-1. communications. However, there is no guarantee that interference will not occur in a The ACTA ID number for the cable modem is particular installation. If this equipment does AU7MMO1BMT5656SMI.
  • Page 37 This device may not interfere with stations operating in the 400.150-406.000 MHz band  in the meteorological aids, meteorological- satellite, and earth exploration-satellite services and must accept any interference received, including interference that may cause undesired operation. CardioMessenger-S TLine Guidelines...
  • Page 38 Warranty Note  The CardioMessenger and all original The CardioMessenger is protected from components by BIOTRONIK are not subject to disturbances resulting from electro- warranty when used improperly or stored and magnetic interference, electrostatic transported incorrectly. Use only the original discharges, and other sources –...
  • Page 39: Technical Data

    +10°C (50°F) to +40°C (104°F) Telephone cable Storage and transport temperature: -10°C (14°F) to + 60°C (140°F) Type: VK-117 Relative humidity: Length: approx. 2.5 m (8.2 ft.) 30% to 75% (non-condensing) Outlet: Rj-11 Atmospheric pressure: 700 hPa to 1060 hPa CardioMessenger-S TLine Technical data...
  • Page 40 100–240 V~ 50–60 Hz 300 mA Output voltage: Output voltage: 5 V DC ± 5%, 2.4 A/12 W 5.1 V DC ± 5%, 2 A/10.2 W Safety class: II Safety class: II Technical manual Technical manual with brief instructions guide CardioMessenger-S TLine Technical data...
  • Page 41: Appendix

     Interference of harmonics according Class A according to IEC 61000-3-2 intended for residential purposes. to IEC 61000-3-2 Voltage fluctuations/flicker  Complies emissions according to  IEC 61000-3-3 CardioMessenger-S TLine Appendix...
  • Page 42 Electromagnetic emitted interference according to IEC 60601-1-2 The CardioMessenger is suitable for operation in the indicated electromagnetic environment.  The customer and/or operator of the CardioMessenger should make sure that it is used in an electro- magnetic environment as described below. Testing ...
  • Page 43 CardioMessenger should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the CardioMessenger. b) Above the frequency range of 150KHz to 80MHz, ensure that field strengths are less than 3V/m. CardioMessenger-S TLine Appendix...
  • Page 44 Electromagnetic emitted interference according to IEC 60601-1-2 The CardioMessenger is suitable for operation in the indicated electromagnetic environment.  The customer and/or operator of the CardioMessenger should make sure that it is used in an electro- magnetic environment as described below. Checking...
  • Page 45 The magnetic field strength should correspond to the typical supply frequencies  level value in business and hospital environments. (50/60 Hz) According to  IEC 61000-4-8 Comment: U is the mains alternating voltage before applying the test levels. CardioMessenger-S TLine Appendix...
  • Page 46 Recommended separation distances between portable and mobile RF communications equipment and the CardioMessenger Transmission  150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz frequency Equation D= 1,17 D= 1,17 D= 2,34 Rated power  Safe distance (m) Safe distance (m) Safe distance (m)
  • Page 47: Index

    DC plug ..... . . 5, 23, 38 BIOTRONIK Service Center ..4, 21 Debugging .
  • Page 48 Home Monitoring ..... 3 Package Contents ....5 Hospital .
  • Page 49 Warranty ......36 Yellow light ....18, 22, 24 CardioMessenger-S TLine Index...
  • Page 52 © BIOTRONIK SE & Co. KG BIOTRONIK SE & Co. KG All rights reserved. Specifications subject to Woermannkehre 1 modification, revision and improvement. 12359 Berlin · Germany Tel +49 (0) 30 68905-0 ® CardioMessenger is a registered trademark of Fax +49 (0) 30 6852804 BIOTRONIK SE &...

Table of Contents