  • Page 1: Monitoring System

    Model: TBG-460GC Blood Glucose / Cholesterol 2 in 1 Monitoring System User's Manual 2012/03/12 Rev.00...
  • Page 2: Table Of Contents

    Table of Contents Chapter 1 TBG-460GC Monitoring System 1.1 Introduction................1.2 Precautions................1.3 Intended Use................1.4 Labelling and Informations............. 1.5 System Components............... Chapter 2 Setup the meter 2.1 Set the Year/Month/Day/Hour/Minute/ Measurement unit..Chapter 3 How to perform Test 3.1 Before Testing................
  • Page 3: Chapter 1 Tbg-460Gc Monitoring System

    Chapter 1 TBG-460GC Monitoring System 1.1 Introduction Thank you for choosing Blood Glucose / Cholesterol Monitoring System. Please read the "User's Manual" for your convenient and for safe use. The Blood Glucose / Cholesterol Monitoring System provides you with accurate, plasma-calibrated results based on electrochemical sense technology and the superior capillary sampling.
  • Page 4: Intended Use

    - Keep the meter and test strips at room temperature (10-30 C). 1.3 Intended Use TBG-460GC Blood Glucose and Cholesterol Monitoring System is intended to be used for quantitative measurement of blood glucose level and cholesterol level in fresh capillary whole blood drawn from the fingertips.
  • Page 5: Labelling And Informations

    1.4 Labelling and Informations Do not re-use Read instructions Caution, consult Keep dry accompanying documents Operation temperature Storage temperature limitation limitation EU Representative In-vitro diagnostic Catalogue number Lot number (Product number) Use by Manufactured by CE certification 0123 1.5 System Components Meter (Front Side) Test Strip Port: Insert test strip here.
  • Page 6 Meter (Back Side) Battery Cover : For holding a 3V lithium battery Tag : Display information about the meter. Glucose Test Strip (Blue Color) Electronic Contact Bar : Sample Inlet : Insert Electronic Contact Bar of the The inlet where the whole blood or control test strip into the Test Port of the solution is drawn in.
  • Page 7: Chapter 2 Setup The Meter

    Code Strip ( Front View ) Contact Bar : Insert this end into Code Number the Test Port of the meter. GLUC 0709 CHOL A412 Test Mode : GLUC - Glucose CHOL - Total Cholesterol Chapter 2 Setup the meter 2.1 Set the Year/ Month/ Day/ Hour/ Minute/ Measurement unit Press and hold the “memory”...
  • Page 8: Chapter 3 How To Perform Test

    3.1 Before Testing Coding the Meter TBG-460GC meter should be coded before testing with test Strips for the first time, or every time a new box of test strips is used. Each vial of test strips has an assigned code number. The code strip is packaged with the test strip in order to calibrate/code the meter.
  • Page 9 Directions : 1. Insert code strip into the test port of the meter. Wait for the code number and test mode to appear on the screen. 设定 2. Make sure the code number on the screen is the same as the code number on the code strip, and the test mode is also the same as marked on the test vial.
  • Page 10: Perform The Blood Glucose Test

    1. Remove the end cap by gently pushing it down and away from the lancing device. 2. Insert the lancet firmly into the lancet holder until it comes to a stop and remove the protective lancet cover. 3. Replace the end cap by snapping it back onto the lancing device body.
  • Page 11: Perform The Total Cholesterol Test

    2. Be careful to make sure the test mode and code number on the screen are the same as the one marked on the test strip vial. Icon 设定 “GLUC” and “CHOL” indicate measuring mode of glucose and total cholesterol respectively. If not, remove the test strip and re-code meter.
  • Page 12: Special Messages

    Note : The TBG-460GC Monitoring System includes glucose and total cholesterol test that are all plasma calibrated. The user does not have to convert it into plasma result. 2. Use the strips before the expiry date, the strips can only be used within 3 months remaining shelf life after the first opening of the...
  • Page 13: Chapter 4 Quality Control

    Chapter 4 Quality Control Intended to use: The control solution is used to check the performance of the system, including Blood Glucose and Cholesterol Monitor System, Glucose and Cholesterol strip, user's test skill. The system is performing adequately if the control solution test result falls within the indicated control range listed on the test strips vial label.
  • Page 14: Chapter 5 Meter Memory Function

    Chapter 5 Meter Memory Function TBG-460GC meter automatically stores up 360 blood glucose and 50 cholesterol test results. It also provides you with 7-, 14-, 21- and 28-day averages of the blood glucose test results. Please follow these directions to recall stored test results.
  • Page 15: Chapter 6 Installing/Replacing The Batteries

    5. After the 28day-average, the latest stored result M01 appears on the screen. Then you can read the stored result from M01 to M360(Glucose mode) or from M01 to M50(Cholesterol mode) by pressing Down or UP Button. 6. Press the “memory” button for over 3 seconds and the meter will turn off after a long “beep”...
  • Page 16: Chapter 7 Storage And Handling Of Meter And Strips

    Chapter 7 Storage and Handling of Meter and Strips Due to sensitive to heat and humidity, the performance of test strips may be affected by improper storage and handling. Please store the test strips and meter carefully according to the specifications listed in chapter 8.
  • Page 17 Message Cause Solution 1. There may be a problem Power off then retry, if the error with self-check of the appears again, contact the meter. authorized distributor for service. 2. Use of the wrong or Check that the code number on damaged code strip or the meter display matches the the code strip was...
  • Page 18: Chapter 9 Specification

    Chapter 9 Specification ± 20% at Glucose level ≥75 mg/dL (4.17mmol/L) System Accuracy ≥ ± 20% at Total Cholesterol 150mg/dL(3.88mmol/L) Principle Electrochemical Biosensor Plasma-equivalent Calibration Test Sample Fresh Capillary whole blood Measuring Time About 10 seconds (Glucose), 26 seconds (Total Cholesterol) Measuring Range Glucose 20 - 600 mg/dL (1.1 - 33.3 mmol/L) Total Cholesterol 100 - 400 mg/dL (2.59 - 10.35 mmol/L)