SAFETY WARNINGS Read First: Safety and Operational Information The international electrical symbols used on the instrument or in this manual are described below. Warning or Caution: Risk of damage or destruction to equipment or software. See explanation in the manual. Warning: Risk of electric shock.
Page 3
Class 1 and Class 2 Laser Products Under no circumstances should a user look directly into optical connectors. Some sources produce invisible radiation that can permanently damage your eyes. Under no circumstances should a user view the device’s optical connectors using magnification unless the device is detached from the SCT series tester.
UNIT FEATURES ....................................3 Physical Features ....................................3 Device laser safety and operation labels .............................4 Cleaning Connectors and Adapters ..............................5 Cleaning SCT-MMA/SMA Fiber Optic and Bulkhead Connectors....................6 Attaching SCT-MMA/SMA Fiber Optic Adapters ..........................6 Multimode Testing Without Mandrels..............................6 Setting a Reference.....................................6 TEST MANAGEMENT ....................................7 PROJECT FOLDERS .....................................7...
Page 5
FIGURES Figure 1 Physical Features ........................................3 Figure 2 Secondary Mode Method B Reference Setting..............................13 Figure 3 Loopback Mode Method B Reference Setting..............................14 Figure 4 Secondary Mode Method B Certification Testing ............................16 Figure 5 Loopback Mode Method B Certification Testing.............................17 Figure 6 Example of Secondary Light Source Mode.............................19 Figure 7 Power Meter Modes ..................................20 Figure 8 Visual Fault Locator ..................................23...
OVERVIEW The SCT-MMA and SCT_SMA fiber optic adapters are used to convert the SCT series of testers into a fully compliant Tier 1 multimode and singlemode fiber optic certification tester. The SCT fiber optic solution offers powerful capability and features including length measurement, two-fiber, dual-wavelength loss measurements, single and bi-directional fiber measurements, power meter mode, light source mode, Fiber Map and visual fault finder (VFL) capability.
Send email to: USTechSupportGrp@megger.com. For US time zone DoverTechnicalSupport@megger.com For UK time zone To order accessories or get the location of the nearest Megger distributor or service center, call: USA: 1 800 723 2861 from USA +1 214 333 3201...
UNIT FEATURES The following sections introduce the fiber optic adapter’s basic features. Physical Features Under no circumstances should a user look directly into the devices optical output connectors. Some sources produce invisible radiation that can permanently damage an eye. Figure 1 Physical Features Visual fault locator mode indication LED.
Cleaning Patch Cord and Fiber Connector End Faces Megger recommends using a cartridge or pocket style dry cleaner to clean fiber optic cable connector end faces. Follow the cartridge or pocket style dry cleaner manufacturer’s directions.
Cleaning SCT-MMA/SMA Fiber Optic and Bulkhead Connectors The best method for keeping the SCT-MMA/SMA and bulkhead connectors clean is to cover the connectors when not in use and only insert clean fiber optic patch cord connectors. Cleaning the SCT-MMA/SMA and bulkhead connectors should always be followed by inspection to determine results.
TEST MANAGEMENT Managing the test environment of the SCT is all organized under the HOME page using the “Manage” functions. These functions allow the user to manage Projects, Cables, Test Limits and Stored Results. PROJECT FOLDERS The Project Folder contains all certification testing parameters, results and information about a specific test site. A project folder is created by selecting the required certification testing parameters and then saving the selections in a project folder.
Fiber optic projects require the following test parameters: Field Description Project Determines the project name. Media The media type selection twisted pair, singlemode fiber or multimode fiber determines what test parameters are required and what adapter is expected. Test Limit The test limit selection determines what tests to perform and how to evaluate the results.
Page 14
Page 1 Select “Projects” from the “HOME” menu using the navigation keys and press [ENTER]. Select “Create Project” and press [ENTER]. Create a Project Name - Navigate to the project name field and press [ENTER]. Using the editor create a new project name. When complete navigate to the Save hyperlink and press [ENTER].
Selecting bi-directional testing will measure and report loss on each fiber at two wavelengths in both directions. When selected the user will be asked to swap fibers on both adapters. Do not disconnect the reference cords from the adaptors, instead disconnect and swap the fibers under test at the first adapters.
Create Cable: Using the navigation keys select “Cables” from the “HOME” menu, under the “Manage” window and press [ENTER]. Select “Create Cable” and press [ENTER]. Select a cable media type and press [ENTER]. Enter the cable parameters then navigate to “Save Cable” and press [ENTER]. Delete Cable: Using the navigation keys select “Cables”...
Establish a fiber optic reference only after the SCT Primary and Secondary units and the SCT-MMA/SMA fiber optic adapters have been on for five minutes. Allow additional time for the adapter to reach ambient temperature if the adapters have been stored above or below ambient temperature.
Secondary Mode Reference Using Test Method B If setting a Method B reference using the Megger provided duplex reference cords use both cords to eliminate removing one cord from the adapter and invalidation the reference. If the fiber optic link uses non-SC style connectors refer to Appendix B for establishing a reference using the modified test method B setup.
Loopback Mode Reference Using Test Method B Setting a Method B reference using the Megger provided duplex reference cord. If the fiber optic link uses non-SC style connectors refer to Appendix B for establishing a reference using the modified test method B setup.
CERTIFYING FIBER OPTIC CABLING Certification requires that the appropriate project is selected and a reference is established prior to testing. The reference measurement is invalidated if the reference patch cord is disconnected from the SCT_MMA/SMA fiber optic adapter. Disconnecting and re-connecting a reference patch cord will alter the optical characteristics resulting in a different reference loss power.
Certifying Duplex Cable Using Test Method B All necessary certification test information is displayed and all required certification tasks are available from the Autotest menu to maximize productivity. If the fiber optic link uses non-SC style connectors refer to Appendix B for certifying using the modified test method B setup.
Certifying Simplex Cable Using Test Method B All necessary certification test information is displayed and all required certification tasks are available from the Autotest menu to maximize productivity. If the fiber optic link uses non-SC style connectors refer to Appendix B for certifying using the modified test method B setup.
INDIVIDUAL FIBER OPTIC CABLING TESTS Individual tests require that the appropriate project is selected and a reference is established with exception of the light source and power meter prior to testing. INDIVIDUAL TEST MENU All individual testing tasks such as selecting a test and starting a test are managed from the Individual Test menu simplifying testing. Using the navigation keys select “Individual Test”...
The light source mode can be enabled on the primary device by selecting light source mode in the individual measurement menu or by pressing the light source enable button on the SCT-MMA or SCT_SMA fiber optic adapter. The light source mode can be enabled on the secondary by pressing the light source enable button on the SCT-MMA or SCT_SMA fiber optic adapter.
Power Meter Mode The power meter mode is enabled on the primary device by selecting power meter in the individual measurement menu. The power meter mode continuously monitors and reports optical power. Simplex SC - SC style fiber optic cord Fibre optic link or spool under test...
FIBER OPTIC CERTIFICATION RESULTS The SCT Autotest carries out all tests required by industry standards. If a test is not applicable to the selected test standard the test will be marked ‘informative’ and will not be evaluated for certification. RESULT SUMMARY The result summary showing all measurements in summary format is displayed first when viewing results.
Page 27
Length Length displays the calculated length of the fiber optic link under test. Simplex (loopback mode) length calculates the total fiber optic link delay from the referenced primary output to the referenced primary input fiber optic connections. Duplex (secondary mode) length calculates ½ of the total fiber optic link delay from the referenced primary output to the referenced secondary input fiber optic connections and the referenced secondary output to the referenced primary input fiber optic connections.
VFL OPERATION The VFL is enabled on SCT-MMA or SCT_SMA fiber optic adapter by pressing the VFL enable button. The VFL output operates independently from the SCT and can be enabled or disabled at any time. The first press of the VFL enable button enables continuous VFL mode as indicated by a constant red VFL mode indication LED.
To clean the SCT-MMA and SCT-SMA fiber optics refer to the section titled Getting Started. STORAGE If you are storing the SCT-MMA and SCT-SMA for an extended period of time ensure that the dust caps are in place and the devices are placed in their original packaging.
ACCESSORIES Product Order Code Accessories included with SCT-MMA Quantity of 2 - 62.5/125 um 2 meter duplex SC reference cords 90000-252 Accessories included with SCT-SMA Quantity of 2 - 9/125 um 2 meter duplex SC reference cords 90000-253...
Fax: +1 610 676 8625 E: VFCustomerService@megger.com Megger operate fully traceable calibration and repair facilities, ensuring your instrument continues to provide the high standard of performance and workmanship you expect. Returning your product to Megger - UK and USA service centres When an instrument requires recalibration, or in the event of a repair being necessary, a Returns Authorisation (RA) number must first be obtained from one of the addresses shown.
CERTIFICATION, COMPLIANCE AND REGULATORY INFORMATION Conforms to relevant European Union directives. Laser Safety Output Port: Class 1 VFL Port; Class 2 Complies with EN60825-1 and EN61010-1 (CE) and CFR21...
The industry fiber optic test standards use different naming conventions when referring to test methods. The three test methods are identical yet the names the different. Table 1 provides the name used by Megger and the equivalent names used by the most common industry standards for the test methods.
APPENDIX B: INDUSTRY STANDARD & APPLICATION TEST METHODS Table 2 defines Test Method B industry standards. Test Method B EN50173 ISO 11801 TIA-568-B Table 2 Test Method B Industry Standards Table 3 defines Test Method B industry applications. Test Method B ATMI FDDI Fibre Channel...
Caution Establish a fiber optic reference only after the SCT Primary and Secondary units and the SCT-MMA/SMA fiber optic adapters have been on for five minutes. Allow additional time for the adapter to reach ambient temperature if the adapters have been stored above or below ambient temperature.
When the reference measurement is complete disconnect the center SC to SC fiber optic reference cord and add the two short duplex patch cords converting SC to the connector style used by the link under test which will produce Test Method B results. Alternate Test Method B will accommodate links with the same or different connectors at each end of the link under test.
Page 37