Contents Foreword Intended use Scanner, computer, monitor and accessory kit straumann CaRes scan Cs2 scanner ® ® Computer box monitor box accessory kit Warnings: Precautionary measures Operating elements and connections of the scanner Description of operating elements Setup and installation...
General terms and Conditions of straumann. straumann products must be used in accordance with their accompanying instructions for use. If you would like additional information on the proper use of straumann products, please contact your local straumann distributor.
Product is part of an overall concept and must be used only in conjunction with the corresponding original components and instruments distributed by straumann usa, llC, its ultimate parent company and all affiliates or subsidiaries of such parent company (“straumann”), except if stated other-...
Please check the packaging for external damage. should you discover such damage, please inform your local straumann representative immediately, as potential complaints will otherwise expire. Please store the transport packaging in case you need to safely transport your scanner.
Computer box the contents of the computer box may vary in different countries. For further information, please contact your local straumann representative. Content of the computer box: p Computer The illustration may differ from the current configuration Monitor box the contents of the monitor box may vary in different coun- tries.
Power cable for monitor p usB cable for scanner – computer p Complete set-up aid kit (set-up aid kit, putty holder and putty) p one copy of software user manual p Installation manual for straumann CaRes scan ® ®...
5. oPeRatInG elements anD ConneCtIons oF tHe sCanneR Description of operating elements Position of indication lights Scanner status Laser status Scanner as illustrated from the front Status of scanner and laser based on indication lights Indication lights and meanings: Status Scanner status LED Laser status LED (left)
Page 10
Description of ports Ports on the back of the scanner (1) USB connection to computer usB connection for scanner control which must be connected to the computer with the supplied serial cable. (2) Port of power supply this port of the scanner must be connected to the power supply with the supplied main power cable.
6. setuP anD InstallatIon the scanner works within a voltage range of 100 – 240 V aC. Arrangement/setup of the system p all plug connections should be easily accessible. Unpacking of the scanner Please remove the individual components (foam parts) while unpacking the scanner.
Setup of the scanner Follow these steps to set up the scanner (refer to the illustration on the right): 1. Release the transportation lock on the right side of the scanner and the two transportation locks on the back before the scanner is turned on. Important: All transportation locks have to be released before operation.
CARES Desktop ® ® as a straumann customer, the default option to set up the my.straumann ® CaRes Desktop is activated. with the pre-setting, the server and client are au- ® tomatically started and guides you immediately to the my.straumann CaRes ®...
8. setuP oF tHe InteRnet ConneCtIon Provided that you did not order a special system con- figuration before delivery, the default setting is routing via DHCP. Configuration of the internet connection to establish an internet connection, it is necessary to connect the network card on the back of the computer with a router via the network cable.
Page 15
5. select Internet Protocol (TCP/IPv4) using the left mouse key and then click on Properties. 6. select both Obtain an IP address automatically and Obtain DNS server address automatically. 7. then click OK. 8. start Internet explorer to test the connection; if the home page is displayed, your computer is connected to the internet.
Configuration of an internet connection with a static IP address 1. log into the windows operating system. 2. Click on the Start button. 3. enter Network connections in the search field and then click on Show network connections. 4. Right-click on Local Area Connection and then select Properties.
Page 17
6. select both Use the following IP address and Use the following DNS server addresses. 7. enter valid IP addresses in both fields. 8. then click OK. 9. start Internet explorer to test the connection; if the home page is displayed, your computer is connected to the internet.
9. tRouBlesHootInG Please refer to the software user manual to find potential causes for errors and their solutions. 10. CHanGInG FIne Fuses If necessary, remove the insert with an appropriate screw- driver and change the corresponding fine fuses (5 x 20 mm, 2.0 at 250 V aC compliant with IeC 60127-2 ss 3).
11. CleanInG notes the following recommendations refer to the scanner, the standard PC housing, the plastic housing, and the lCD monitor with surface made of coated and/or painted plastic. Cleaning of scanner p use a damp towel to clean the outside of the scanner. Please make sure that the scanner is unplugged while doing so.
Page 20
Dust in the interior of the computer If you suspect dust in the scanner, PC housing, plastic housing, or the lCD monitor do not open. Contact straumann technical support. opening the housing of any hardware will void the warranty. Caution Do not open the housing while the computer is connected to the power supply.
Foam parts for transportation It is necessary to use the original packaging components for the safe transportation of the scanner. Straumann CARES Scan CS2 – Packaging overview ® ® transportation box for the straumann CaRes scan Cs2 ® ® scanner...
Page 22
Foam components number 1 to 3 (2 for each side; left and right) and component number 4 (one piece for the top). Please note that the actual pieces are not numbered and that this is an illustration only. Foam components for the bottom of the box (numbers 5 –...
Page 23
2. Insert component 6 into component 5. Place component 7 on the bottom of the box as illustrated. 4. Insert components 5 – 6 (as illustrated in step 2) as protection on the back of the box.
Page 24
5. Carefully place the scanner into the box. Cover the scanner with a protective plastic wrap. 7. Position components 1–3 (prefabricated as described in step 1).
Page 25
8. Place component number 4 across as shown in the illustration on the right. 9. Close the lid of the box and tape it shut.
13. teCHnICal Data Technical data model designation: CaRes scan Cs2 ® Connections: usB 2.0 High-speed/power supply Power: 100 – 240 V aC 50 – 60 Hz Power Input max.: 900 ma laser protection class: Class 1, 1m IP protection: IP20 –...
14. exPlanatIon oF symBols Protective earth laser radiation LASER RADIATION laser warning DO NOT VIEW DIRECTLY WITH OPTICAL INSTRUMENTS CLASS 1M LASER PRODUCT Caution – Risk of electrical shock serial number order code Fuse Consult instructions for use Do not dispose of with household waste Caution: Federal law restricts this device to sale by or on the order of a dental professional.