Digital Multimeter
Function Generator
10 Digit Counter
MODEL : 81K / 81K-TRMS
80000 Series
1
Brief introduction
These series have two models.
The 80000 is autorange and calibrated by 50Hz sine signal.
The 80000R is also antorange and can offer true RMS measurement.
Safety information
Read this operation manual completely before using the meter to ensure that you use the meter safety
follow the safety guidelines as below :
Use the meter only as specified in this manual; Other wise,the protection provided by the meter
l
maybe impaired.
l
Never measure voltage while the test leads are inserted into the current input terminals.
l
Do not use the meter if it looks damaged .
l
Inspect the leads for damaged insulation or exposed metal, check test lead continuity. Replace
damaged leads.
l
Disconnect the power and discharge all high-voltage capacitors before testing in resistance, continuity,
and diode function.
l
Be Cautions when working above DC 60V or AC 42V,such voltages may cause a shock hazard.
l
When making measurement, keep your fingers behind the guards plant on the probes.
l
Select the proper function and range for measurement, To avoid damaging the meter, disconnect
the test leads from test points before change function.
3
A unique Multi-functional device
®
Features :
l
Multi-Display : Primary 80000
Secondary 80000
Bargraph 23 segments
l
Basic DC Accuracy : 0.05%
l
mV Impedance : >1000MV
l
More than 50 measuring functions.
l
Frequency measurement : 0.5Hz to 1000.000000 MHz
l
16 Freguency points, 1% to 99% duty selectable square wave output.
l
Analyzing component of current or voltage signal.
Resistance : 0.1V to 80MV,10MV to 8000MV
l
l
Capacitance : 1PF to 100mF
Temperature : -50
0
C to 1372
l
l
dBm 20 types of reference impedance.
Auto data hold/peak hold.
l
l
Relative measurement.
l
36-hours dynamic recording : MAX / MIN / AVG / MAX-MIN
l
Communication : isolated optical RS232
l
Timer for measurement.
l
Backlight display.
l
Auto power off
Function of meter
Terminal
About terminal function refer to table 1
Table 1. terminal
Terminal
COM
V / V / Hz
mA
10A
0
C, -58
0
F to 2502
0
F
Function
Common terminal for all measurement
Volts, Ohm, Diode, Freq., Temp. and Cap.
measurement and square wave output terminal
milliampere current measurement terminal
Ampere current measurement terminal
Figure 1. terminal
2
4

Advertisement

loading
Need help?

Need help?

Do you have a question about the 81K and is the answer not in the manual?

Questions and answers

Summary of Contents for Meco 81K

  • Page 1 Relative measurement. 36-hours dynamic recording : MAX / MIN / AVG / MAX-MIN 10 Digit Counter Communication : isolated optical RS232 Timer for measurement. MODEL : 81K / 81K-TRMS Backlight display. 80000 Series Auto power off Brief introduction Function of meter Terminal These series have two models.
  • Page 2 Rotary switch 1. SELECT About rotary switch function refer to table 2 Press this button you can select your measurement mode. Table 2. functions of rotary switch When meter as square wave output, press this button can select duty of square wave (1% - 99%). position of Function rotary switch...
  • Page 3 present readings as Relative reference value and display on secondary. Relative measurement has two Display mode. one is RELr = measuring value - Reference value, the other is REL %= (RELr /Reference value) Table 4. symbols of display x 100% ( press SELECT to select RELr or REL% mode). Order No.
  • Page 4 3. After setting proceeding measurement. In the auto range (Auto) mode, the meter selects the best range for the input detected. This allows you to switch test points without having to reset the range. a. If measuring above limited the secondary display “ ”...
  • Page 5 AC/DC millivoltage measurement interference. Some random digits on display but have not effect The measurement of millivoltage has three modes : DCmV / [ACmV+Hz] / dBm. on result of measurement. 1. Set the rotary switch to “ mV ” position. 3.
  • Page 6 5. Turn off power to the circuit, discharge all high-voltage capacitors. 3. Touch the probes to the test points and reading display. 6. Break the circuit path to be tested. Touch the black test lead to more negative side of the break. Touch Note : The test leads can add 0.1 V ~ 0.5 V of error to resistance measurement, please short test leads the red test lead to the more positive side of the break.
  • Page 7 Temperature measurement Diode and continuity check 1. Set the rotary switch to “TEMP” position. Caution 2. Press SELECT to select Hi or normal test mode. To avoid possible damage to the meter or to the equipment under test, disconnect circuit power and discharge all high-voltage capacitors before checking diodes.
  • Page 8 Table 13. ACV AVG (60Hz Sine Wave Calibrating) Table 16. ACA AVG (60Hz Sine Wave Calibrating ) Range Resolution Accuracy Range Resolution Accuracy Note 50Hz / 60Hz 50Hz - 1kHz 1kHz - 10kHz 10kHz - 20kHz 80mV 80mV ±(1.0%0 rdg + 8) ±(1.5% rdg + 8) ±(4.0% rdg + 8) ±(8.0% rdg + 8)
  • Page 9: Certificate Of Calibration

    This meter has a communication capability. This function will assist user to recording and keeping data easy. We have affer RS232 MECO METERS PVT. LTD. ADAPTER to optional accessories. The RS232 ADAPTER include a cable with photoelectric receiver and a software disc. We will Block 9, Plot 270, 2nd Floor, Rup-Udey Niwas, Sion (E), Mumbai - 400 022 (INDIA) continuous release ncw version per year.