User Manual for PC-80D Easy ECG Monitor
Instructions to User
Dear Users,
Thank you very much for purchasing our product. Please read the manual very carefully before using this device.
These instructions describe the operating procedures to be followed strictly. Failure to follow these instructions may cause
measuring abnormality, device damage, or personal injury.
No part of this manual may be photocopied, reproduced or translated into another language without the prior written consent.
Creative reserves the right to improve and amend it at any time without prior notice.
Version of This Manual: Ver1.1
All rights reserved.
3502-2430011
I

Advertisement

Table of Contents
loading

Summary of Contents for Delton PC-80D

  • Page 1 User Manual for PC-80D Easy ECG Monitor Instructions to User Dear Users, Thank you very much for purchasing our product. Please read the manual very carefully before using this device. These instructions describe the operating procedures to be followed strictly. Failure to follow these instructions may cause measuring abnormality, device damage, or personal injury.
  • Page 2 User Manual for PC-80D Easy ECG Monitor WARMING 1. Check the device to make sure that there is no visible damage that may affect user’s safety and measurement performance. Stop using the unit, when there is obvious damage. 2. Stop using the ECG cable if it is damaged, please change a new one.
  • Page 3: Table Of Contents

    User Manual for PC-80D Easy ECG Monitor Table of contents 1 Overview ..................................... 1 1.1 Appearance .................................... 1 1.2 Name and Model ................................... 2 1.3 Structure ....................................2 1.4 Features ....................................2 1.5 Intended Use ..................................3 1.6 Operating Environment ................................. 3 2 Installation ....................................
  • Page 4 User Manual for PC-80D Easy ECG Monitor 5 Waveform Review ..................................12 5.1 Record list ................................... 12 5.2 Waveform Review ................................13 6 Data Management ..................................15 6.1 Storage status ..................................15 6.2 Delete ID ..................................... 15 6.3 Delete all ..................................... 16 6.4 Upload Data..................................
  • Page 5 User Manual for PC-80D Easy ECG Monitor 10.4 Display ....................................23 10.5 Data Memory ..................................23 10.6 Overall Dimension and Weight ............................23 10.7 Additional declaration ................................ 23 11 Maintenance .................................... 24 12. Troubleshooting ..................................25 13 Key of Symbols ..................................26 Appendix 1 Additional measuring method ..........................
  • Page 6: Overview

    User Manual for PC-80D Easy ECG Monitor 1 Overview 1.1 Appearance Power: Long time press (about 2 seconds) makes the device power on or off; short time press turns on/off the backlight. Back: back to upper level of menu. 3. Metal Electrode I Measurement: shortcut button for built-in electrodes measuring, press this button to start measuring.
  • Page 7: Name And Model

    Name: Easy ECG Monitor Model: PC-80D 1.3 Structure PC-80D Easy ECG Monitor mainly consists of circuit boards, display screen and metal electrodes/lead wires. 1.4 Features 1. Small in size, light in weight, and easy to carry. 2. One-key measurement, easy to operate.
  • Page 8: Intended Use

    User Manual for PC-80D Easy ECG Monitor 1.5 Intended Use This Easy ECG Monitor is intended for measuring and recording theECG waveform and average heart rate of adult patient. It is applicable for use in clinics and homes, and convenient to operate by the patients themselves.
  • Page 9: Default Menu Screen

    User Manual for PC-80D Easy ECG Monitor 3 Default menu screen 3.1 Menu screen ”Power key (lasting for 2 seconds) until a beep sound is heard to indicate the unit is turned on, then the booting screen will Long time press the“...
  • Page 10: Ecg Measurements

    User Manual for PC-80D Easy ECG Monitor 4 ECG measurements There are 2 measuring modes for the ECG measurement, which includes quick measurement by built-in electrodes and long term measurement by external lead wires. Quick measurement: press “ ” measure key to enter into the quick measurement directly; it can display single channel ECG signal measured by built-in electrodes, the signal can be lead I, lead II or chest lead according to the electrode placement.
  • Page 11 User Manual for PC-80D Easy ECG Monitor Note: Keep at ease and make sure the skin contact well with the electrodes during ECG measurement. Please place the device on correct position, do not shake it during the measurement. Dirty skin may cause poor signal quality, so clean your skin before measuring.
  • Page 12 User Manual for PC-80D Easy ECG Monitor Operation: Connecting the lead wires after placing the electrodes well and then insert the connector onto the device; Power on the device to select the “limb lead” for signal input and set the measuring time on option setup menu, then press “”ok key to start the measurement.
  • Page 13: Measuring Procedure

    User Manual for PC-80D Easy ECG Monitor 4.2 Measuring procedure 4.2.1 Preparing Phase The quick measurement is as following for details: At default menu screen, press “ ” measure key to enter into the quick measuring screen directly, as shown in figure 4-7, in addition, at the beginning of measurement (including quick and long term measurement), the preparing phase shows on the screen with a heart beat icon and counting down seconds, the beeper will sound with the counting down at the interval of one second.
  • Page 14: Chest Lead Measurement

    User Manual for PC-80D Easy ECG Monitor 4.3.1 Chest lead measurement Please refer to the section 4.1 “Chest lead measurement” to connect the lead wires; enter “ ” option setup menu to select “chest lead” for signal input; return to default menu screen, move the cursor to the “...
  • Page 15 User Manual for PC-80D Easy ECG Monitor (9) “ ” :Status bar of measurement progress. Operation: When in the quick measurement by built-in electrodes:  ① Press “ ” key, the measuring display screen will be made a quarter turn counterclockwise;...
  • Page 16: Data Analysis And Measurement Result Description

    User Manual for PC-80D Easy ECG Monitor 4.4Data analysis and measurement Result Description When the measuring is finished, the device will analyze the measurement result, as shown in figure 4-11/4-12/4-13; the measurement result description refers to the index 2. Figure 4-12(no irregularity found in Ext.)
  • Page 17: Waveform Review

    User Manual for PC-80D Easy ECG Monitor 5 Waveform Review 5.1 Record list Select “ ” review key on the default menu screen, then press “” ok key to enter into the user ID list screen, as shown in Figure 5-1. Selecting a folder which contains the data for a given ID,then press OK key,the display screen is as shown in Figure 5-2...
  • Page 18: Waveform Review

    User Manual for PC-80D Easy ECG Monitor 5.2 Waveform Review In the record list, selecting a record, then press “”ok key, the quick measurement record is shown in the figure 5-3; the external lead wire measuring records is shown in figure 5-4.
  • Page 19 User Manual for PC-80D Easy ECG Monitor In the figure 5-7/5-8/5-9 screen, press the “” ok key can review the result description for this ID record, as 00:00:01 X1 shown in figure 5-10/5-11. In the measure result screen, press the “” ok key can delete this record. Note: ①...
  • Page 20: Data Management

    User Manual for PC-80D Easy ECG Monitor 6 Data Management Select “Memory” icon on the default menu, then press “ OK” button to enter into data management screen, as shown in Figure 6-1. 6.1 Storage status On data management screen, select “Statistic”, then press “ ” ok key to enter into the Storage Figure 6-1 status screen, as shown in Figure 6-2.
  • Page 21: Delete All

    When measuring is finished, on data management screen, select “Upload data” to enter into upload data screen, as shown in figure 6-5. On this screen, you can transmit data to PC for data management, and interpretation through PC-80D data cable. Please refer to the ECG Viewer Manager for detailed operation.
  • Page 22: Measurement Option Setting

    User Manual for PC-80D Easy ECG Monitor 7 Measurement option setting On the default menu,select the “Option”,then press the “”ok key to enter into the optional parameters setting for measurement, as shown in figure 7-1. On-screen display: User ID: Displaying the established ID No. or “New ID”. At most 100 ID can be created. When creating a new ID, at most 30 characters can be entered;...
  • Page 23: Creating New Id

    User Manual for PC-80D Easy ECG Monitor 7.1Creating new ID On the “Option" setup menu, selecting “ New ID” item from the ID pull-down menu, as shown in figure7-2. Operation: 1. Press the navigation key “,,or”: Move the cursor upwards/backwards.
  • Page 24: System Setup

    User Manual for PC-80D Easy ECG Monitor 8 System Setup On the default menu screen, select “Setting” with Navigation Key, then press “”ok key button to enter into system setup screen, as shown in Figure 8-1. 1.Press the navigation key “,,or”:Move the cursor upwards/backwards.
  • Page 25: Auto Power Off

    User Manual for PC-80D Easy ECG Monitor 8.4Auto power off The timeout setting for auto power off is shown in the figure 8-5. Press navigation key“,,or” to set the timeout time. Press the “ ” back button when setting is done.
  • Page 26: Help

    User Manual for PC-80D Easy ECG Monitor 9 Help On default menu screen, select “Help”, then press “” ok key to enter into help info screen. In the info screen,you can browse the “FAQ”, “How to”, “ECG info” and software version information. As shown in 9-1.
  • Page 27: Technical Specifications

    User Manual for PC-80D Easy ECG Monitor 10 Technical Specifications 10.1 ECG measurement 1. Amplification channels: 3. 2. Signal input: 3 embedded metal electrodes or external lead wire cable (with 3/4 lead wires). 3. Lead selections and measuring methods: Quick measurement by embedded electrodes; measurement by external electrodes with 3 lead wires (single channel for Lead I, II, or III) or 4 lead wires (3 simultaneous channels for Lead I, II and III) 4.
  • Page 28: Classification

    User Manual for PC-80D Easy ECG Monitor 10.3 Classification 1. The type of protection against electric shock: Internally powered equipment 2. The degree of protection against electric shock: Type BF applied part 3. The degree of protection against harmful ingress of liquid: Ordinary equipment without protection against ingress of liquid.
  • Page 29: Maintenance

    User Manual for PC-80D Easy ECG Monitor 11 Maintenance 1. Please change the batteries, when appears low-voltage icon. 2. It is recommended to wipe the metal electrode with medical alcohol, when the measuring signal is too weak. 3. Please wipe the surface of the unit with clean cloth or let it dry in air.
  • Page 30: Troubleshooting

    User Manual for PC-80D Easy ECG Monitor 12. Troubleshooting Trouble The device can not turn on. Possible Reason Solution 1. The batteries are drained or almost drained. 1. Change batteries. 2. The batteries are not inserted 2. Reinstall batteries. properly.
  • Page 31: Key Of Symbols

    User Manual for PC-80D Easy ECG Monitor 13 Key of Symbols Symbol Description Symbol Description Heart Rate (Unit: beats per minutes) Battery indicator Power button shortcut button for ECG measuring ▇ Confirmation button Back button ,,, Navigation key USB Port...
  • Page 32: Appendix 1 Additional Measuring Method

    User Manual for PC-80D Easy ECG Monitor Appendix 1 Additional measuring method Direction of using limb clip type electrodes: clean the electrodes by using wet cloth; plug the lead wires into socket; or affix the lead wires to clip, and then keep the clip in measuring position.
  • Page 33 User Manual for PC-80D Easy ECG Monitor 4. F(green)electrode location of Chest lead measurement The location of ECG electrodes: electrode location for V1-V6 on the 4 intercostal space at right border of sternum on the 4 intercostal space at left border of sternum...
  • Page 34: Appendix2 Measurement Result Description

    User Manual for PC-80D Easy ECG Monitor Appendix2 Measurement Result Description No. ECG Measurement Result Descriptions Notes: (bpm:time/minute) No irregularity found. 60 bpm ≤ HR ≤ 100 bpm, No irregular HR found. Suspected a little fast beat 100 bpm < HR <= 110 bpm Suspected fast beat HR>...
  • Page 35 User Manual for PC-80D Easy ECG Monitor Suspected a little slow and interval with A little slow beat and suspected Premature beat. occasional short beat interval Suspected a little slow beat and interval A little slow beat and suspected Bigeminy /Trigeminy.
  • Page 36: Appendix3 General Knowledge Of Ecg

    User Manual for PC-80D Easy ECG Monitor Appendix3 General Knowledge of ECG ⅠNormal sinus rhythm: In sinus conditions, SA node paces the heart with the regularly rate and the normal rhythm. The heart rate is in the range of 60 to 100 times per minute and the rhythm is regular. P wave is normal and each one is followed by a QRS wave.
  • Page 37 User Manual for PC-80D Easy ECG Monitor Indication: It occurs when healthy people fall asleep, and it can be found in athlete (or those who love doing sports frequently), old people, or vagus excitement person. Pathology: Sick sinus syndrome, Ischemic heart disease, Cardiomyopathy, intracranial hypertension, increased hypokalemia, Low temperature, period of convalescence of acute infectious disease or after use some medicines such as digitalis.
  • Page 38 User Manual for PC-80D Easy ECG Monitor 3) Bigeminy: It is a type of PVC in which a normal beat is coupled with a premature beat. Because of its trigger cardiac abnormal pacemaker in different position, it divided into arial, ventricular and node.
  • Page 39 User Manual for PC-80D Easy ECG Monitor Suggestion:If it occurs time after time, please see the doctor as soon as possible. 2) Paroxysmal ventricular tachycardia:Caused by Premature Ventricular Contraction, Heart Rate >140bpm. Indication:Most commonly found in heart disease patient, it can cause ventricle fibrillation if it’s serious, so the tester needs to see a doctor immediately.