Special Calibration - Kipp & Zonen CT24 Instruction Manual

Signal amplifier 4-20 ma
Hide thumbs Also See for CT24:
Table of Contents

Advertisement

5.2

Special calibration.

With special calibration is meant that the sensitivity of the sensor
(pyranometer) is used to adjust the gain. This is to obtain for
instance a conversion of 200 W/m
Example:
CM 11 with a sensitivity of 5.14 x 10
irradiance of 2000 W/m
The zero point is set as described above under standard
calibration.
2
For 200 W/m
/ mA the applied voltage to the input to set the gain,
is calculated in the following way:
-6
5.14 x 10
x 2000 = 10,28 mV, and adjust the output to
+4 mA + (2000/200) = 14 mA
This should always be smaller than 20 mA.
An input signal change of 0 to 2000 W/m
of 4 to 14 mA.
With the 500 Ohm resistor this results in a voltage change of 2 to 7
Volt.
Depending on the sensor output it can be better to use a
conversion of 50 or 100 W/m
Of course using the total +16 mA range gives the optimal
resolution on the output.
20
INSTRUCTION MANUAL CT 24
CALIBRATION PROCEDURE
2
per mA.
-6
V / W/m
2
2
gives an output change
2
per mA.
2
and max.

Advertisement

Table of Contents
loading

Table of Contents