B-2. Product Description B-2-1. ELITE 600 Power Supply Hardware Overview C. Installation of ELITE 600 Power Supply ---------------------------------- 5 C-1. Package List C-2. Setup and Operation of the ELITE 600 Power Supply C-2-1. Unpacking C-2-2. Setup and operation D. Troubleshooting ------------------------------------------------------------------ 8 D-1.
Wealtec Corp. A-1. Warranty ELITE 600 are warranted to be free from defects in materials or workmanship for a period of one year from the original invoice date, under normal usage. Any defects occurring during warranty period, Wealtec Corp. will repair or replace defective products or parts without charge unless the defects arise from conditions outlined below.
Be sure that there is at least 6 cm clearance around the ELITE 600 power supply to ensure adequate cooling. Always connect the power supply using the AC power cord provided with the power supply or a similar 3-prong, grounded AC outlet.
ELITE 600 V1.0 B. Introduction ELITE 600 power supply is compact, space saving and light weight in design. It can be used with many electrophoresis applications, such as running standard PAGE gels, multiple PAGE mini gels, horizontal/vertical agarose gels, semidry transfer and minigel tank transfer. The LED display reports power supply status.
Starts and Stops the output from the Power Supply The front and rear panel buttons as shown above are used to set up and run the ELITE 600 power supply. The audible alarm design allows a beep sound for each and every button pressed on the front membrane keypad.
Unwrap all packages carefully and compare contents with the packing list, ensure the arrival of all the items. If any part is missing, contact your local Wealtec representative. Inspect all the components for damage that may have occurred while the unit was in transit. If any part appears damaged, contact the carrier immediately.
Page 8
ELITE 600 V1.0 1. Connect the main power cord: Connect the main power cord to the power cord receptacle on the rear panel and to a suitable grounded AC power outlet. 2. Connect the HV power leads: Connect the apparatus to the power supply by plugging the lead connectors into one pair of the recessed output jacks.
Page 9
ELITE 600 V1.0 PFD: The Power Failure Detection feature will save the time, voltage and current settings in the event of a power outage and automatically continue the run when the power is restored. To alert the user that a power failure occurred during the run, the error message “E6” flashes at the end of the run instead of the usual “OFF”...
LED & Display do not light up fuse if necessary and try again Front panel power on indicator off 3. Check if power switch malfunction 4. Contact your local Wealtec representative Fan is spinning 1. Check main power source LED & Display do not light up 2.
ELITE 600 V1.0 D-2. Error codes The ELITE 600 power supply has built-in error messages to help troubleshoot problems with the electrical circuit. The error codes and possible corrective actions are listed below: Error Explanation Possible Solution Code No load detected 1.
ELITE 600 V1.0 1. Check for any short circuit or evidence of electrical arching. Short Circuit 2. Buffer concentration is too high. Load current exceeded 500 mA 3. Press any button* to clear the error code. *Press any button, except the RUN button, to clear the error message.
E-mail: support@wealtec.com Wealtec Corp. offers complete technical support for all of our products. If you have any questions about how to use this product, or would like to arrange to repair it, please contact your local Wealtec Corp. representative or international technical / service specialist by...
1001006 ELITE 300 Plus* Power Supply, 230 V, 50~60 Hz 1001011 ELITE 600* Power Supply, 100~240V, 50~60 Hz 1001007 Mini ELITE* Power Supply, 90~264 V, 50~60 Hz * Standard package: Complete power supply unit, an instruction manual, power cable and a...