I N T R A O R A L D I G I T A L X - R A Y I M A G I N G S Y S T E M U S B I n t e r f a c e M o d e l Part Numbers: B5100 and B5200 I n s t r u c t i o n Manual...
A C C E N T USB Intraoral Digital X-ray Imaging System. References in this manual to Air Techniques includes its division AllPro Imaging. Refer to the following companion documents as necessary:...
Congratulations on your purchase of the A C C E N T USB Intraoral Digital X-ray Imaging System, the latest imaging product from Air Techniques, a leading manufacturer of dental, medical and veterinary equipment since 1962. The A C C E N T...
S A F E T Y N O T I C E This equipment has been designed to minimize exposure of personnel to hazards. While the A C C E N T is designed for safe operation, certain precautions must be observed. Use of the A C C E N T not in conformance with the instructions speci- fied in this manual may result in permanent failure of the unit.
IEC 601-1 and/or IEC 601-1-1 harmonized national standard. Use of ACCESSORIES or cables other than those specified or provided by Air Techniques' may result in increased EMISSIONS or decreased IMMUNITY of the EQUIPMENT. Electrical Safety Notes.
P P / / N N B B 5 5 1 1 0 0 0 0 Size 1 Sensor B5120 Accent Sensor Care Guide B5028 Disposable Infection Control Sheaths for Size 1 B5125 Sensor (pack of 25)
U N P A C K I N G A N D I N S P E C T I O N Unpack each component of the ACCENT and inspect for physical damage such as scratched panels, damaged connectors, etc. If any damage is noted, notify your authorized dealer immediately so corrective action can be taken.
K The operator is liable for all damages caused as a result of incorrect usage of this appliance. K Only A C C E N T Sensors made by Air Techniques may be used with the A C C E N T Control Module.
T E C H N I C A L D A T A Power Supply The USB A C C E N T Control Module is a USB powered device using less than 450 mA. No external power supply is required. Operating Conditions Temperature 50 to 105°F (10 to 40°C)
Page 11
T E C H N I C A L D A T A A C C E N T Sensor Size 2 Number of pixels 2.05 million Pixel Size 22 x 22 μm Technology CCD + Scintillator Weight 3.4 ounces. (95 g) Dimensions Length Width...
To operate the A C C E N T Sensor, it must be connected to a Important: compliant Computer System, and the computer must be loaded with authorized Image Management Software. Air Techniques provides neither the PC nor the Software. The PC and Software are provided by the dealer or user.
The A C C E N T can only be used when connected to a user-supplied computer system loaded with an Image Management Software authorized by Air Techniques. Each time the A C C E N T Sensor is used it must be covered using an...
INDICATORS AND CONNECTORS Front Panel. The specific functions of the LED indicators and connector (see Figure 2) located on the front panel of the Control Module are described below. INDICATORS GREEN BLUE YELLOW 20-PIN SENSOR CONNECTOR SOCKET Figure 2. USB A C C E N T Control Module Front View Green LED indicator Illuminates green to indicate that the connection between the Control...
The TWAIN Test application displays the Accent TWAIN user interface providing controls to the break existing wireless link or establish a new one. The Accent TWAIN user interface ‘remembers’ the last link that was used, so when you return to your imaging application, acquisition will use the new settings.
Page 16
TWAIN INTERFACE Figure 4. TWAIN Graphical User Interface 1. Title Bar: Contains the words "Accent: TWAIN Data Source". If it displays any other words, then you are using a language other than English or you have not selected the Accent as your TWAIN data source.
I N S T A L L A T I O N Protect the imaging system from direct sunlight and excessive heat. Ensure that the Control Module connector is accessible at all times for connection of the A C C E N T sensor as necessary plug Important:...
Page 18
I N S T A L L A T I O N A C C E N T USB System Interconnection. Refer to Figure 5 and perform the following procedures to connect the A C C E N T system. 1.
If the Control Module locks up, reset by unplugging the USB cable and then plugging back in. Initial Setup This procedure installs Accent TWAIN Interface into Windows. It must be performed once on each computer that will use the Accent Control Module. Perform the procedures as follows: 1.
Page 20
. . y y y y y y y y y y is selected, where x and yyyyy are num- bers representing the version of the Accent TWAIN interface Press the S S e e l l e e c c t t software.
When the TWAIN interface for the ACCENT sensor is open, the yellow LED should flash. If it does not, you must verify that your ACCENT sys- tem is powered on, has a sensor plugged in and no other software is try- ing to communicate with the ACCENT system.
Page 22
OPERATION Observe that the yellow LED blinks to indicate the system is ready to take a new image. When the X-ray machine is activated, the Control Module auto- matically detects the exposure signal and stores the image in internal memory. At this point, the yellow LED extinguishes and the blue LED illuminates.
OPERATIONAL OPTIONS Image Resolution The ACCENT system allows image acqusition in one of two resolutions, High resolution and Low resolution.Each has distinct advantages and disadvantages that will have to be taken into account when making choice for your particulat application.
C A R E A N D M A I N T E N A N C E Sensor Care To prevent damage to the A C C E N T sensors, please read and act on the following suggestions wipe down the sensor with a sterile solution store the sensor in its proper location when it is not in use keep the sensor and the remote off of the floor at all times...
Page 25
The ACCENT Sensor can be cleaned using a disinfection spray or wipe such as Birex® se. To avoid malfunction of the ACCENT Sensor, make sure that the plug of the sensor does not come in contact with the cleaning agent.
T R O U B L E S H O O T I N G Trouble Possible Cause Corrective Action 1. Green LED not lit. • Trouble with USB Power • Make sure the computer supply to Control Module. power is on. •...
ACCESSORIES The following lists the ordering number and description for accessory components available to maintain and expand the ACCENT Sensor to meet your professional needs. Contact your authorized dealer for information. I I t t e e m m P P a a r r t t N N o o . .
Page 28
Air Techniques and AllPro Imaging are leading manufacturers of fine dental, medical and veterinary equipment from air and vacuum systems and X-ray film processors, to an impressive line of new products incorporating the most recent technological advances. These new products, vital...