1. Blood Glucose Measurement
What happened?
The monitor does not display a message after inserting a test strip.
POSSIBLE CAUSE
Batteries exhausted
Batteries incorrectly
installed or absent .
Test strip inserted upside
down or not completely
inserted.
Defective monito r.
What happened?
The test does not start after applying the sample.
POSSIBLE CAUSE
Insu fficient blood sample.
Defective test strip.
Sample was applied
when flashing blood Symbol
was not displayed
.
Defective monito r.
What happened?
If you turned off the monitor but it makes a beeping sound.
POSSIBLE CAUSE
The strip is still inside the slot .
44

Problem in Operation

WH AT TO DO
Replace the batteries.
Check that the batteries ar e
correctly installed.
Insert the test strip correctly
with the contact bars end first
and facing up.
Contact local customer service
WH AT TO DO
Repeat the test with a new test
strip and a larger sample.
Repeat the test with a new
test strip.
Repeat the test with a new test
strip. Apply sample only when
appears on the display .
Contact local customer service
WH AT TO DO
R emove the strip. If you need to
test blood glucose, insert an
unused strip.
flashing