Function check

432
25 If you think your NovoPen 3 PenMate is not working properly, follow this procedure:
433
a) Check that the barrel and the NovoPen 3 PenMate are screwed together tightly and
434
that the dose indicator window on the barrel is aligned with the yellow push button on
435
the NovoPen 3 PenMate.
436
b) Screw on a new NovoFine needle as described in step10.
437
c) Clear the air bubbles as described in step11.
438
d) Put the outer needle cap over the needle.
439
e) Dispense 20 units into the needle cap.
440
441
The insulin should fill the lower part of the outer needle cap. If the pen has delivered
442
too much or too little insulin, repeat the test. If it happens again, do not use the pen.
443
Contact your supplier.
444
Do not try to repair a faulty NovoPen 3 or a faulty NovoPen 3 PenMate.
445
Never use NovoPen 3 or NovoPen 3 PenMate unless you are sure that they are
446
working properly.