kENDALL KANGAROO ePump Operation And Service Manual

kENDALL KANGAROO ePump Operation And Service Manual

Enteral feeding pump
Table of Contents

Advertisement

tyco/ Healthcare
kENDALL
KANGAROO
Enteral Feeding Pump
Operation and Service Manual
Enteral Feeding Pump — Operation and Service Manual
Nutripompe — Manuel d'utilisation et d'entretien
Pumpe für die enterale Ernährung — Betriebs- und Serviceanleitung
Pompa di alimentazione enterale — Manuale operativo e di manutenzione
Bomba para alimentación enteral — Manual de operación y servicio
Enteral matningspump — Användar- och servicemanual
Enterale voedingspomp — Handleiding voor bediening en onderhoud
Bomba de alimentação entérica — Manual de funcionamento e assistência
Enteraalinen ravitsemuspumppu — Käyttö- ja huolto-opas
Pumpe til enteral ernæring — Betjenings- og servicehåndbog
Αντλία εντερικής σίτισης — Εγχειρίδιο λειτουργίας και σέρβις
Pumpa pro enterální výživu — Provozní a servisní příručka
Enterális táplálópumpa — Kezelési és karbantartási kézikönyv
Насос для энтерального питания — Руководство по эксплуатации и техническому обслуживанию
Pompa żywienia dojelitowego — Instrukcja obsługi i serwisowania
Enteral Besleme Pompası — Operasyon ve Servis Kılavuzu
Enteral ernæringspumpe — Bruker- og vedlikeholdshåndbok
Rx
ONLY
0123
See Instructions
™ Trademark of Tyco Healthcare Group LP or its affi liate.
™*SPRAY-NINE is a trademark of Spray-Nine Corporation.
rademark of Spray-Nine Corporation.
rademark of
™*PHISOPHEX is a trademark of Smith Kline Beecham Corp.
rademark of Smith Kline Beecham Corp.
rademark of
™*HIBICLENS is a trademark of Tubifoam Ltd.
rademark of Tubifoam Ltd.
rademark of
™*VESTA-SYDE is a trademark of Steris, In
rademark of Steris, In
rademark of
TYCO HEALTHCARE GROUP LP • MANSFIELD, MA 02048
MADE IN U.S.A. • © 2005 TYCO HEALTHCARE GROUP LP
ALL RIGHTS RESERVED • 050328
PRODUCT INFORMATION IN U.S. 1-800-962-9888 • www.TycoHealthcare.com
EC REP: TYCO HEALTHCARE U.K. LTD • GOSPORT, PO13 0AS, U.K.
SWISS DISTRIBUTOR: TYCO HEALTHCARE SWITZERLAND LTD.
ROOSTRASSE 53, POSTFACH CH-8832, WOLLERAU SWITZERLAND
182388
c.
e
Pump

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the KANGAROO ePump and is the answer not in the manual?

Questions and answers

john tindall
July 2, 2025

what does KTO mean on pump

Georges Bourgeois
April 18, 2025

Bonjour: J'ai besoin de changer de batterie, le No est C21355035 07-2021. Dans ma recherche je ne trouve pas ce no. Ce que j'ai trouvé, donne le No OEM F010484. J'aimerais savoir quel No commander. Merci.

Georges Bourgeois
April 18, 2025

covodien Kangaroo ePump

Subscribe to Our Youtube Channel

Summary of Contents for kENDALL KANGAROO ePump

  • Page 1 Healthcare kENDALL Pump KANGAROO ™ Enteral Feeding Pump Operation and Service Manual Enteral Feeding Pump — Operation and Service Manual Nutripompe — Manuel d’utilisation et d’entretien Pumpe für die enterale Ernährung — Betriebs- und Serviceanleitung Pompa di alimentazione enterale — Manuale operativo e di manutenzione Bomba para alimentación enteral —...
  • Page 2: Table Of Contents

    © 2005 Tyco Healthcare Group LP English Thank you for purchasing the KANGAROO Pump™ Enteral Feeding Pump. With proper care, this device will provide you with years of precision service. TABLE OF CONTENTS Page Section I — General Information ..................5 Section II —...
  • Page 3: Enteral Feeding Pump

    © 2005 Tyco Healthcare Group LP English TABLE OF CONTENTS Page Continuous Mode (Adjust Feed) .........................16 Continuous Mode Feeding ........................16 Continuous Mode Flushing ........................17 Intermittent Mode (Adjust Bolus) ......................18 Intermittent Mode Feeding ........................18 Intermittent Mode Feeding - Bolus Max ....................Bolus Max ....................
  • Page 4 © 2005 Tyco Healthcare Group LP English TABLE OF CONTENTS Page Section IX — Error/Warning/Informational Screens and Troubleshooting System Error ..............................29 Hold Error ..............................30 Rotor Error ..............................31 Feed Error ..............................31 Flush Error ..............................31 Flow Error ..............................32 Pump Set Dislodged Error ..........................32 Battery Low ...............................33 Feeding Complete .............................33 Pump Set Use >24 Hours Warning ......................33...
  • Page 5: Section I - General Information

    English Section I – General Information The KANGAROO ePump enteral feeding pump is a simple-to-use, precision enteral feeding pump. It is easily programmed to provide patients with either continuous or intermittent feeding, and can provide automatic flushing capability when used with KANGAROO ePump Feed & Flush Sets.
  • Page 6 © 2005 Tyco Healthcare Group LP English ����������� ���� ���������� ����������� �������������� ���� ��������� ������� ������������� ������������ ������� ������� �������������� ����� ����������������� ��������������� ������������ ���������� ���������������� ���������������� ��������������� ���� ������� ����������� ���� Figure 1A. KANGAROO ePump features, front and rear views. Pump KANGAROO ™...
  • Page 7: Section Ii - Safety And Warnings

    4. Use only KANGAROO ePump enteral feeding pump Power Cord (with built-in A/C adapter) with pump. See Section XIII - Service Part Numbers for replacement of Power Cord. 5. Use only KANGAROO ePump enteral feeding pump Pump Sets with this device. Pump is not compatible with other Pump Sets.
  • Page 8: Section Iii - Icon Identification

    © 2005 Tyco Healthcare Group LP English 11. See icon descriptions in Section III-Icon Identification and Section X - Specifications and Symbols for additional safety information. 12. This device is designed to minimize the effects of uncontrolled electromagnetic interference and other types of interference from external sources.
  • Page 9: Section Iv - Initial Setup

    © 2005 Tyco Healthcare Group LP English Error Indicator Pump operations are stopped until the error is resolved. Information Indicator This indicates information only, and does not require any immediate action. Section IV — Initial Setup Attaching the A/C Adapter Power Cord Plug the Power Cord into the A/C Adapter Port on the back of the pump.
  • Page 10: Attaching Pole Clamp

    To attach the Pole Clamp to the KANGAROO ePump enteral feeding pump, align the hole on the Pole Clamp with the mounting hole on the back of the pump. Use a 3/8” - 16 bolt to fasten together, as shown in Figure 3.
  • Page 11: Section V - Instructions For Use

    © 2005 Tyco Healthcare Group LP English Section V — Instructions For Use Quick Start 1. Press POWER on lower, right-hand corner of control panel. 2. Open blue door enclosing Pump Set loading area. 3. To load Pump Set (refer to Figure 4): •...
  • Page 12: General Startup

    • Placed feet down on any horizontal surface. A/C Power Operation Plug the KANGAROO ePump enteral feeding pump into an A/C power outlet for normal operation. When the pump is not plugged in, or if A/C power is interrupted, a built-in rechargeable battery will run the pump.
  • Page 13: Loading Pump Sets

    English Loading Pump Sets The KANGAROO ePump enteral feeding pump will indicate LOAD A SET in the screen’s status line if a Pump Set is not LOAD A SET in the screen’s status line if a Pump Set is not LOAD A SET installed.
  • Page 14: Prime Pump

    Prime Pump The KANGAROO ePump enteral feeding pump may be primed automatically with a single button push, including the flushing line if a Feed & Flush Set is loaded. The pump can also be primed in a more interactive method using the hold- to-prime feature.
  • Page 15: Auto Priming

    © 2005 Tyco Healthcare Group LP English Auto Priming The AUTO PRIME option will be available if the pump senses that the currently installed Pump Set was not previously primed, and that there is no fluid in the line near the rotor. If auto-priming is desired, press and release4AUTO PRIME to automatically prime the line(s).
  • Page 16: Selecting Continuous Or Intermittent Feed Mode

    English Selecting Continuous or Intermittent Feed Mode The KANGAROO ePump enteral feeding pump can be programmed to feed continuously or intermittently. The continuous feeding mode will deliver the enteral nutrition at a steady rate, either until the programmed volume has been delivered or until the supply has been exhausted.
  • Page 17: Continuous Mode Flushing

    The pump will automatically limit flushing settings which exceed pump capabilities to deliver. CAUTION: The KANGAROO ePump enteral feeding pump flushes at a rate of 1960 mL/hr (32.7 mL/minute). Use care when programming the flush volume so that it matches the patient’s need.
  • Page 18: Intermittent Mode (Adjust Bolus)

    © 2005 Tyco Healthcare Group LP English Intermittent Mode (Adjust Bolus) Intermittent Mode Feeding For INTERMITTENT feeding mode, the ADJUST BOLUS option is available in the opening menus (Figure 7C for a Feed-Only set or Figure 7D for a Feed & Flush set). Select this option to program the feeding parameters from the ADJUST BOLUS screen, Figure 12A.
  • Page 19: Intermittent Mode Feeding - Bolus Max

    © 2005 Tyco Healthcare Group LP English Figure 12A. ADJUST BOLUS menu. Figure 12B. Setting the Bolus Rate. Figure 12C. Setting the Number of Boluses. Figure 12D. Setting the volume per Figure 12E. Setting the time interval Bolus. between starts of bolus. Intermittent Mode Feeding –...
  • Page 20: Run

    The KANGAROO ePump enteral feeding pump will only run when a Pump Set is properly loaded and the necessary feeding parameters have been programmed. For a Feed & Flush Set, the flushing parameters do not have to be set for the pump to begin operating.
  • Page 21: Hold

    © 2005 Tyco Healthcare Group LP English The RUN option can be selected from the following screens, if the proper parameters are programmed and the Pump Set is loaded: • SET LOADED screens (Figure 7) SET LOADED screens (Figure 7) SET LOADED •...
  • Page 22: Flush Now

    © 2005 Tyco Healthcare Group LP English Clear Volume Fed Press4CLEAR VOLUME to re-zero the output feed totals “mL FED, ” “BOLUSES FED, ” and “mL FLUSH. ” Adjust Settings Press 4ADJUST SETTINGS to return to one of the opening menus shown in Figure 7, thus allowing any setting to be modified.
  • Page 23: More Options

    © 2005 Tyco Healthcare Group LP English More Options Press4MORE in one of the opening menus, Figures 5 or 7, to access the MORE OPTIONS screen, Figure 16. MORE OPTIONS screen, Figure 16. MORE OPTIONS Figure 16. MORE OPTIONS menu. OPTIONS menu.
  • Page 24: Biotech Menu

    © 2005 Tyco Healthcare Group LP English BioTech Menu The BioTech screen will show technical information about the pump, and has a lock feature to lock the Pump Settings against unauthorized changes. Accessing the BioTech Screen To get the BioTech screen, press the power button to start the pump, and press and hold the top4button while the startup sequence (Kangaroo icon “hops”...
  • Page 25: Lock Screen Feature

    Lock Screen Feature The KANGAROO ePump enteral feeding pump can be set to RUN and then locked out from unauthorized use. While the LOCK SETTINGS feature prevents only changes to the input parameters, the LOCK SCREEN feature will lock out any button presses, including RUN or HOLD, except power off.
  • Page 26: Certifying The Accuracy Of Pump Flow Rate

    If any of the following events occur, DO NOT USE the pump until it has been properly cleaned and serviced by personnel trained in servicing KANGAROO ePump enteral feeding pumps: • wetting of the pump’s power cord or leakage into the pump interior during cleaning.
  • Page 27 English General Cleaning Directions Cleaning of KANGAROO ePump enteral feeding pumps must be performed as follows: CAUTION: Disconnect pump from AC power source before cleaning. After cleaning, do not connect to AC power source until pump and power cord are thoroughly dry.
  • Page 28: Section Viii - Battery Replacement

    • Push connector in until properly latched. • Tuck wires into slot between connector and battery to insure wires are not pinched. Re-tighten Battery Door screw. Note: Replacement Battery Packs are available from Tyco Healthcare/Kendall Co. (See Section XIII - Service Part Numbers).
  • Page 29: Section Ix - Error/Warning/Informational Screens And Troubleshooting

    English Section IX — Error/Warning/Informational Screens and Troubleshooting For operational errors, the KANGAROO ePump enteral feeding pump display will show the error indicator icon, with specific information regarding the error condition, and will sound the alarm. The following conditions will trigger error conditions and will activate audible and visible alarms: •...
  • Page 30: Hold Error

    © 2005 Tyco Healthcare Group LP English Error # Related Sub-system Time Out during downstream occlusion check, never got RotorOn3=0. May blow fuse. Could not find Stop Cock Flush Position during prime routine. Could not find Stop Cock Feed Position during prime routine. Motor Error during Tubing AutoLoad.
  • Page 31: Rotor Error

    © 2005 Tyco Healthcare Group LP English Rotor Error The ROTOR ERROR screen appears during RUNNING or RUNNING or RUNNING PRIMING, when the pump has detected an unusual operating condition for the rotor. See Figure 24. The ROTOR ERROR generally results from a problem with the Pump Set tubing around the rotor.
  • Page 32: Flow Error

    © 2005 Tyco Healthcare Group LP English POWER to stop operation of the pump and put a different pump into If the error still cannot be resolved, press If the error still cannot be resolved, press If the error still cannot be resolved, press service.
  • Page 33: Battery Low

    © 2005 Tyco Healthcare Group LP English Battery Low The BATTERY LOW screen appears and the alarm beeps continuously when the battery needs to be recharged. There is approximately 15 minutes of battery life remaining when this screen appears. Plug the power cord in to begin recharging. The pump will automatically return to the screen that was active prior to the error.
  • Page 34: Led Array

    Type Infusion Device Volumetric Pumping Mechanism Rotary Peristaltic Pump Sets KANGAROO ePump MISTIC Feed-Only Set or Feed & Flush Set Feeding Formula Delivery Rate 1-400 ml/hr in 1 ml increments Feeding Formula VTBD 1-3000 ml in 1 ml increments Bolus Volume...
  • Page 35 © 2005 Tyco Healthcare Group LP English Flushing Solution Dose Range 10-500 ml in 1 ml increments Flushing Solution Interval Range 1-24 hr in 1 hr increments Accuracy ±10% or 0.5 ml/hr, whichever is larger, with bag at 46 cm (18”) above pump, at room temperature 22° C ± 2° C (72° F ±...
  • Page 36: Symbols On Pump And Accessories

    � � 2. Grasp black ring retainer and gently stretch tubing around rotor Alternating Current Do not use Kangaroo ePump � 3. Pull up on retainer and insert into Feeding Sets for greater than right pocket Class II Equipment (degree of...
  • Page 37: Section Xi - Customer Service

    Certain replacement items, as listed in Section XIII - Service Part Numbers, are available from the service centers listed below. All service personnel must be properly trained and qualified with operation of the KANGAROO ePump enteral feeding pump. Improper service may impair operation of the pump.
  • Page 38: Battery Pack

    Visit our web site at: www.TycoHealthcare.com (1) KANGAROO ePump ....(US) 382400 ......(International) 482400 (2) Main Door (Blue transparent door) .
  • Page 39: Section Xv - Electromagnetic Conformity Declaration

    The KANGAROO ePump enteral feeding pump has been built and tested according to UL60601-1, E181931 CSA-C22.2 No.601.1-M90, and EN60601-1-2 Standards. The KANGAROO ePump enteral feeding pump is intended for use in the electromagnetic environment specifi ed below. The user of the pump should assure that it is used in such an environment.
  • Page 40: Section Xvi - Accusystem

    English Section XVI – AccuSystem√ The KANGAROO ePump is a reliable and accurate enteral feeding device. The reason for the accuracy of the KANGAROO ePump enteral feeding pump is the design, controlled calibration methods, and stringent manufacturing controls. One of the most critical functional parts of the KANGAROO ePump enteral feeding platform is the AccuSystem√ feeding set produced by Tyco Healthcare.
  • Page 41 Pump KANGAROO ™ Performance Certifi cation KANGAROO ePump™ Enteral Feeding Pump Serial Number Date of Re-Certifi cation The undersigned hereby certifi es that the above referenced pump has been tested per the pump’s internal re-certifi cation protocol with the following results:...
  • Page 42 ™ Enteral Feeding Pump Performance Certifi cation Operation and Service Manual KANGAROO ePump™ Enteral Feeding Pump Enteral Feeding Pump — Operation and Service Manual Nutripompe — Manuel d’utilisation et d’entretien Serial Number Pumpe für die enterale Ernährung — Betriebs- und Serviceanleitung Pompa di alimentazione enterale —...