Maintenance and Testing

Cleaning the Lancing Device

Clean the outside of the ACCU-CHEK Multiclix lancing device regularly with 70% isopropyl (rubbing) alcohol.
DO NOT place the entire device under water. DO NOT use bleach. At least once a week, disinfect the cap after
cleaning by placing it in 70% rubbing alcohol for 10 minutes. Allow the cap to air-dry after disinfecting.
Chapter 5: Maintenance and Troubleshooting
Chapter 1: Your New Meter
The meter needs little or no maintenance with normal use. It automatically tests its
own systems every time you turn it on and lets you know if something is wrong.
See Chapter 5, "Screen Messages and Troubleshooting."
If you drop the meter or think it is not giving accurate results, make sure that your
test strips and control solution haven't expired, then run a control test.
Test your display before each blood sugar test to make sure it is working properly.
To test the display, turn the meter off, then press and hold
display. All the segments should be clear and look exactly like the picture to the left.
If one of the segments is missing or looks different from the picture, do not use the
meter. Call the ACCU-CHEK Customer Care Service Center at 1-800-858-8072.