Do you have a question about the PURELAB Option-R 7 and is the answer not in the manual?
Questions and answers
KHIRI Redha
May 10, 2025
bonjour,
J'ai un problème avec cet appareil, le problème est qu'elle ne démarre pas. J'ai vérifié la partie électrique, tout est bon : l'alimentation 220V et 24V, et les fusibles 6 fusibles, tout est bon. Mais la machine, elle ne s'allume pas.
Salutation, merci d'avance.
1 comments:
Mr. Anderson
May 20, 2025
If the ELGA Option-R 7 machine does not start despite having a good electrical supply and functioning fuses, possible reasons could include:
1. The mains power is not switched on. 2. The electrical connections are loose or not secured. 3. The feed water supply is not connected or is inadequate. 4. The reservoir is empty or not properly connected. 5. Operating temperature is outside the 1 to 35°C range, which can damage the unit. 6. The inlet water pressure is either too low (below 4.0 bar) or too high (above 6 bar) without a pressure regulator.
Each of these conditions should be checked and corrected as needed.
Cartridge(s) ............28 SANITIZATION PROCEDURES........29 Sanitization of Unit and Reservoir....... 31 CT1 Sanitization Tablet - Safety Information ..33 TROUBLE SHOOTING..........34 CONSUMABLES AND ACCESSORIES ..... 35 KEY TO CONTROL PANEL........36 Page ii PURELAB Option-R 7/15 Version 4 08/07...
Customer Support PURELAB Option-R Service support and consumable items are available from your local supplier or distributor. Refer to customer service contact details shown at the end of this publication. PURELAB Option-R 7/15 Version 4 08/07 Page 1...
Control of Substances Hazardous to Health (COSHH) Material safety data sheets covering the various replaceable cartridges are available upon request. Contact your local supplier or distributor. UV Lamp Page 2 PURELAB Option-R 7/15 Version 4 08/07...
(See Section 10 - Consumables and Accessories, for detail). Dispense tap Sanitization port Power switch ON/OFF Mains power socket Removable cover Control Panel Fuse Alternative dispense Feedwater inlet tap position connection Door Removeable cover PURELAB Option-R PURELAB Option-R 7/15 Version 4 08/07 Page 3...
0.2μm bacterial filter can be fitted to the dispense tap for added protection. • During periods of non-use the unit will automatically operate in intermittent re-circulation mode to maintain water purity with maximum efficiency. Page 4 PURELAB Option-R 7/15 Version 4 08/07...
Page 10
Portable water Purified water inlet Inlet outlet Inlet Temp sensor Sanitization Port Flow Strainer controller Ion-exchange Cartridge pack Solenoid LC141 Pre-treatment LC140 LC105 Concentrate Drain Boost pump (optional) Process Flow - PURELAB Option-R PURELAB Option-R 7/15 Version 4 08/07 Page 5...
Wall, bench or under bench Wall, bench or under bench mounted. mounted. Environment Clean dry indoor. Clean dry indoor. Temp 5 - 40°C. Temp 5 - 40°C. Humidity max 80% Humidity max 80% non-condensing. non-condensing. Page 6 PURELAB Option-R 7/15 Version 4 08/07...
Page 12
RS232 Remote display connection Volt free contact-internal Safety Features Power fail safe Boost pump protection from particulates Re-circulation pump protection from particulates Low operating voltage 24V Audio visual alarms Adjustable alarm settings PURELAB Option-R 7/15 Version 4 08/07 Page 7...
Page 13
Refer to flow tables outside these conditions. Subject to correct operating and maintenance procedures. As part of our policy of continual improvement we reserve the right to alter the specifications given in this document. Page 8 PURELAB Option-R 7/15 Version 4 08/07...
Page 14
PURELAB Option-R Reverse Osmosis Capacity Charts Option-R7 Pressure (bar) Graph 1 - Nominal Flowrate vs Inlet Pressure for PURELAB Option-R7 Option-R15 Pressure (bar) Graph 2 - Nominal Flowrate vs Inlet Pressure for PURELAB Option-R15 PURELAB Option-R 7/15 Version 4 08/07 Page 9...
Right hand The PURELAB Option-R control panel has a range of control icons control button control button as follows: Control Panel Button Icon Description LEFT Menu Scroll RIGHT Reset Mute Alarm Accept Printer Page 10 PURELAB Option-R 7/15 Version 4 08/07...
• Temperature Range 5 - 40°C • Maximum Relative Humidity 80% @ 31°C decreasing linearly to 50% @ 40°C, non-condensating The unit is in Installation Category II, Pollution Degree 2, as per IEC1010-1. Unit Rear Mounting Points PURELAB Option-R 7/15 Version 4 08/07 Page 11...
Page 17
Strainer Feed Water Strainer PURELAB Option-R RESERVOIR Level control cable Pump Inlet feed All tube connections 8mm (5/16") Drain PURELAB Option-R Unit Installed with Storage Reservoir Page 12 PURELAB Option-R 7/15 Version 4 08/07...
(LA512). Connector Collet Blanking transit plug Pull plug Depress collet Clean square Feedwater Inlet Connections cut end Tube 8mm OD Simply push in tube to attach Tube in secured position Fitting Tubes PURELAB Option-R 7/15 Version 4 08/07 Page 13...
Page 19
Electrical Connections Step 3 - Connect High/Low Level Switch to Reservoir 1. INSERT jack plug into the level control socket located at rear of unit and reservoir. Level control Reservoir Level Connections Page 14 PURELAB Option-R 7/15 Version 4 08/07...
Page 20
13. Switch water supply on. Replace side panel 14. Press the process button. 15. Ensure correct operation of dispense tap. Manifold Panel screws fixing screws Replacing side panel Positioning the Dispense Tap PURELAB Option-R 7/15 Version 4 08/07 Page 15...
Select the option required using the button and accept with the button. Audible Alarm Enable/Disable Page 16 PURELAB Option-R 7/15 Version 4 08/07...
Page 22
4380 hours (6 months). Press to jump to the next consumable or to initiate reset. Press to reset timer or press to abort reset. Press to jump to the next consumable. Ion-exchange Cartridge Timer PURELAB Option-R 7/15 Version 4 08/07 Page 17...
Page 23
• UV tube replacement timer (hours remaining). • Pre-treatment cartridge replacement (hours remaining). Step 5 - Report Printing (Only if connected to LA618) To print a report press the icon. Water Conductivity Page 18 PURELAB Option-R 7/15 Version 4 08/07...
13. REMOVE the by-pass block. (Store in the front door). Removing & Inserting By-pass Block 14. INSERT Ion-exchange cartridge LC141. (See Section 7.2). 15. PRESS the PROCESS button to begin the water purification process. PURELAB Option-R 7/15 Version 4 08/07 Page 19...
1 hour. The unit will re-circulate the reservoir contents for 10 minutes every hour. Pressing the PROCESS button at any time during this mode will initiate re-circulation. Flash Normal Process Screens Page 20 PURELAB Option-R 7/15 Version 4 08/07...
Press the button to mute the alarm. If water purity stays outside acceptable limits replace the ion-exchange pack following the instructions in section 7.2. Flash Water Purity Alarm / Standby PURELAB Option-R 7/15 Version 4 08/07 Page 21...
Page 27
This alarm condition will cause the process to turn off. Connect the Reservoir Level reservoir level and press the process key to clear the alarm. Disconnect Alarm Page 22 PURELAB Option-R 7/15 Version 4 08/07...
Standard conditions are 4 bar (60 psi) inlet pressure at 15°C, potable water with clean pre-filter. CAUTION! Ensure that the display and replacement timer settings are reset after replacing consumables. (Refer to section 5.4). PURELAB Option-R 7/15 Version 4 08/07 Page 23...
4. REFIT the outlet tubing at the top of the cartridge. 5. RESET pre-treatment cartridge timer. (See Section Setting Display Replacement Timer). 6. PRESS the PROCESS button to start the unit. 7. CHECK the unit for leaks and close front door. Page 24 PURELAB Option-R 7/15 Version 4 08/07...
7. RESET ion-exchange pack timer. (See section 5.4 - Setting up Display and Replacement Timer). Replacing Ion-exchange 8. PRESS the process button to start the unit. Cartridge 9. CHECK the unit for leaks. PURELAB Option-R 7/15 Version 4 08/07 Page 25...
7. PUSH UV unit into the retaining clips. 8. CLOSE the front door. 9. RESET UV alarm settings. (See Section 5.4 - Setting Up Display and Replacement Timer). 10. PRESS the PROCESS button to start the unit. LC105 Lamp Page 26 PURELAB Option-R 7/15 Version 4 08/07...
1. REPOSITION the inlet strainer. 2. REFIT tubes to inlet strainer, ENSURE it is facing the correct direction. 3. REPOSITION the pre-treatment cartridge into its support clips. 4. RE-ESTABLISH inlet water supply. 5. TURN on power. PURELAB Option-R 7/15 Version 4 08/07 Page 27...
Customer Service. WARNING! ALL NEW RO CARTRIDGES ARE FILLED WITH A BACTERIOSTATIC SOLUTION TO PREVENT BACTERIAL CONTAMINATION DURING STORAGE. THE RO CARTRIDGES WILL THEREFORE REQUIRE RINSING PRIOR TO USE. Page 28 PURELAB Option-R 7/15 Version 4 08/07...
CT1 Cleaning Tablet 2. UNSCREW cap on sanitization by-pass block. 3. INSERT CT1 tablet. 4. REFIT cap on by-pass block, hand tight. By-pass Block PURELAB Option-R 7/15 Version 4 08/07 Page 29...
Page 35
Step 9 - Return to Normal Operation 1. TURN on power. 2. RECONNECT the outlet tube to the reservoir. 3. PRESS PROCESS button to return to normal operation. 4. CHECK the system for leaks. Page 30 PURELAB Option-R 7/15 Version 4 08/07...
By-pass Block 2. PRESS the PROCESS button to start the sanitization process. 3. ALLOW the sanitization PROCESS to continue for 60 minutes. 4. PRESS the PROCESS button to stop the sanitization process. PURELAB Option-R 7/15 Version 4 08/07 Page 31...
Page 37
Step 10 - Return to Normal Operation Replace Ion Exchange Cartridge 1. TURN on power. 2. RECONNECT the outlet tube to the reservoir. 3. PRESS PROCESS button to return to normal operation. 4. CHECK the system for leaks. Page 32 PURELAB Option-R 7/15 Version 4 08/07...
Mouth - wash out the mouth thoroughly with water and give large quantity of milk to drink. Obtain medical advice. Note: comprehensive Data Sheet available on request and is supplied with each packet of tablets. PURELAB Option-R 7/15 Version 4 08/07 Page 33...
Replace UV Lamp. (See Section 7.3 - Replacing the Ultraviolet Lamp). Unit noisy Open front door and secure pipework to stop vibration. No flow from dispense Push PROCESS button. Check reservoir level is >40%. Page 34 PURELAB Option-R 7/15 Version 4 08/07...
ELGA PURELAB Option-R 7/15 Operator Manual 11.4 Low Level, Quality and Standby Alarms Screen Description Low level alarm Process on and water purity alarm Standby and purity alarm Standby position PURELAB Option-R 7/15 Version 4 08/07 Page 37...
PURELAB Option-R 7/15 Operator Manual ELGA 12. WARRANTY/CONDITIONS OF SALE ELGA LabWater is a trading name of VWS (UK) Ltd. General Limited Warranty VWS (UK) Ltd. warrants the products manufactured by it against defects in materials and workmanship when used in accordance with applicable instructions for a period of one year from the date of shipment for the products.
Page 44
VWS (UK) Ltd. warrants its products against defects in materials and workmanship as described in the Warranty statement on the preceding pages. PURELAB Option-R 7/15 Version 4 08/07 Page 39...
Need help?
Do you have a question about the PURELAB Option-R 7 and is the answer not in the manual?
Questions and answers
bonjour, J'ai un problème avec cet appareil, le problème est qu'elle ne démarre pas. J'ai vérifié la partie électrique, tout est bon : l'alimentation 220V et 24V, et les fusibles 6 fusibles, tout est bon. Mais la machine, elle ne s'allume pas. Salutation, merci d'avance.
If the ELGA Option-R 7 machine does not start despite having a good electrical supply and functioning fuses, possible reasons could include:
1. The mains power is not switched on.
2. The electrical connections are loose or not secured.
3. The feed water supply is not connected or is inadequate.
4. The reservoir is empty or not properly connected.
5. Operating temperature is outside the 1 to 35°C range, which can damage the unit.
6. The inlet water pressure is either too low (below 4.0 bar) or too high (above 6 bar) without a pressure regulator.
Each of these conditions should be checked and corrected as needed.
This answer is automatically generated