SECTION
2
Operation
2.1 INTRODUCTION
This section provides information
needed to use the Model 7154 2-Pole
High Voltage Card with the Models 705 and 706 scanner mainframes.
Once the card is installed in the mainframe,
refer to the mainframe in-
struction
manual for complete operating
instructions.
2.2 SAFETY
PRECAUTIONS
WARNING
User supplied
lethal voltages
may be present on the PC
board
or the connections.
Turn off all Dower and
discharge
stored energy in external circuitry
before mak-
ing or breaking
connections.
1. Do not exceed the Model ?l54 maximum
voltages of -tllOOV peak
terminal
to terminal
and +llOOV peak any terminal to chassis.
2. Make sure the scanner mainframe
is grounded
through
an earth
grounded
receptacle before operation.
3. inspect all test lead connections
for wear and defects such as cracks
and exposed wires.
4. Use appropriately
rated cables when switching
high voltages.
2.3
WIRING
Each channel on the Model il54 card consists of a double-pole.
nor-
mally open (2.Form A) switching
relay. The Model 7154 will switch any
one of the 10 signals (inputs) to one output,
or switch one signal to
any one of 10 outputs.
See Figure 1-l.